{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and  this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.",
      "address_2": "",
      "product_quantity": "699",
      "code_info": "3070039: S/N 2271 to 2793  3074486: S/N 1812 to 3036  3070013: S/N 3967 to 5042  3070021: S/N 3852 to 4719",
      "center_classification_date": "20180817",
      "distribution_pattern": "US Nationwide Distribution to accounts in IL, FL, and NY.",
      "state": "PA",
      "product_description": "3D TOP Ceiling Stand, model no. 3070039  3074486  3070013  3070021    Product Usage:  The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.",
      "report_date": "20180829",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-2837-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80735",
      "termination_date": "20191004",
      "more_code_info": "",
      "recall_initiation_date": "20180629",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}