  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Louis",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80174",
      "recalling_firm": "LIEBEL-FLARSHEIM COMPANY LLC",
      "address_1": "1034 S Brentwood Blvd",
      "address_2": "",
      "postal_code": "63117-1223",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to: AL, AZ, CA, CT, CO, FL GA, HI,  IA, IL, ID,  KS, KY, MD, MI, Michigan, MN, NC, NH, NV, NY, OK, OH, PA, SC, TN, TX, VA,  WA, WI,  and WV,",
      "recall_number": "Z-2834-2018",
      "product_description": "Liebel-Flarsheim Direct Digital Imaging System (DDIS)     Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.",
      "product_quantity": "28",
      "reason_for_recall": "Software issue.  The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%.  The display in all other modes deviated by less than 35% from the dosimeter readings.",
      "recall_initiation_date": "20170227",
      "center_classification_date": "20180821",
      "termination_date": "20230816",
      "report_date": "20180829",
      "code_info": "Product ID  750736  SHF-635RF,   750737 SHF-835RF",
      "more_code_info": ""
    }
  ]
}