  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Kloten",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86085",
      "recalling_firm": "NOBEL BIOCARE SERVICES AG",
      "address_1": "Balz-zimmermannstrasse 7",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of  Albania and Hong Kong.",
      "recall_number": "Z-2828-2020",
      "product_description": "NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.",
      "product_quantity": "188",
      "reason_for_recall": "Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.",
      "recall_initiation_date": "20200720",
      "center_classification_date": "20200820",
      "termination_date": "20210512",
      "report_date": "20200826",
      "code_info": "Lot: 13084072,  UDI: (01)07332747073584(10)13084072(17)240525"
    }
  ]
}