  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wilmington",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71983",
      "recalling_firm": "Ferno-Washington Inc",
      "address_1": "70 Weil Way",
      "address_2": "N/A",
      "postal_code": "45177-9371",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.",
      "recall_number": "Z-2826-2015",
      "product_description": "Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System    Product Usage:  Stretcher for pre-hospital patient transport.",
      "product_quantity": "72 units",
      "reason_for_recall": "Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.",
      "recall_initiation_date": "20150803",
      "center_classification_date": "20150930",
      "termination_date": "20160120",
      "report_date": "20151007",
      "code_info": "Serial #\u0019s: 15F001703, 15F001704, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001715, 15F001717, 15F001719, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001728, 15F001729, 15F001730, 15F001731, 15F001732, 15F001733, 15F001734, 15F001735, 15F001736, 15F001737, 15F001738, 15F001739, 15F001740, 15F001741, 15F001742, 15F001744, 15F001745, 15F001746, 15F001747, 15F001748, 15F001749, 15F001750, 15F001751, 15F001752, 15F001753, 15F001754, 15F001755, 15F001756, 15F001757, 15F001758, 15F001759, 15F001760, 15F001761, 15F001762, 15F001763, 15F001764, 15F001765, 15F001766, 15F001767, 15F001768, 15F001769, , 5F001770, 15F001771, 15F001772, 15F001773, 15F001774, 15F001775, 15F001776, 15F001777, 15F001778, 15F001779, 15F001780, 15F001781, 15F001782, 15F001783, 15F001784, 15F001785, 15F001786, 15F001787, 15F001788, 15F001789, 15F001790, 15F001791, 15F001793, 15F001794, 15F001795, 15F001796, 15F001797, 15F001798, 15F001804, 15F001805, 15F001806, 15F001807, 15F001808, 15F001809, 15F001810, 15F001811, 15F001812, 15F001813, 15F001814, 15F001815, 15F001816 & 15F001850."
    }
  ]
}