{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bedford",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99115",
      "recalling_firm": "Instrumentation Laboratory",
      "address_1": "180 Hartwell Rd",
      "address_2": "N/A",
      "postal_code": "01730-2443",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2825-2026",
      "product_description": "Brand Name: GEM Premier 5000  Product Name: GEM Premier 5000  Model/Catalog Number: 00024019255   Portable blood analyzer system",
      "product_quantity": "21",
      "reason_for_recall": "Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.",
      "recall_initiation_date": "20260508",
      "center_classification_date": "20260729",
      "report_date": "20260805",
      "code_info": "GEM Premier 5000:  Part number: 00024019255  UDI: 08426950870227  Units distributed in Spain:    GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710.  Affected Accessory:  Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026.  Part No. 00024015859; UDI No. 08426950869634"
    }
  ]
}