  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Louisville",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72183",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "826 Coal Creek Cir",
      "address_2": "N/A",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to  Canada; United Arab Emirates; Australia; Belgium; Switzerland; China; Germany; Denmark; Spain; Finland; France; Great Britain; Greece; India; Israel; Italy; Japan; Netherlands; New Zealand; Poland; Portugal; Sweden; Singapore; Thailand; Slovakia; Turkey; and South Africa.",
      "recall_number": "Z-2824-2015",
      "product_description": "P/N: 9734856, SCREWDRIVER, 9734856, SOLERA STD MAST; P/N: 9734857, SCREWDRIVER, 9734857, SOLERA RDN MAST; P/N:  9734279, SCREWDRIVER 9734279 SOLERA STANDARD; and P/N: 9734373, SCREWDRIVER 9734373 SOLERA REDUCTION.  The Medtronic Navigated Solera Screwdrivers are reusable devices that are designed for use in spine surgery to assist in the placement of screws during the surgery, using a Medtronic computer-assisted surgery system.",
      "product_quantity": "2346",
      "reason_for_recall": "Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.",
      "recall_initiation_date": "20150914",
      "center_classification_date": "20150930",
      "termination_date": "20151203",
      "report_date": "20151007",
      "code_info": "For this action, device lot numbers are not applicable.  The action covers all lots of products distributed between September 2010 and July 2015"
    }
  ]
}