  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Williston",
      "state": "VT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86158",
      "recalling_firm": "89 NORTH",
      "address_1": "20 Winter Sport Ln",
      "address_2": "N/A",
      "postal_code": "05495-8144",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-2820-2020",
      "product_description": "LDI Laser firmware lower than V02.13.00 for   LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6    Laser Science fluorescence imaging",
      "product_quantity": "188",
      "reason_for_recall": "The defect relates to firmware version lower than V02.13.00. It is controlling  safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - \"Interlock Override Mode\" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).",
      "recall_initiation_date": "20200131",
      "center_classification_date": "20200825",
      "termination_date": "20230502",
      "report_date": "20200902",
      "code_info": "LDl-7, LDI-WF, LDI-NIR, LDl-6"
    }
  ]
}