  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86187",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "1450 E Brooks Rd",
      "address_2": "N/A",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US states:   CA  CO  CT  GA  HI  IL  IN  KY  MA  MD  NC  NY  OK  PA  SD  TN  WA  WI    OUS countries:   AE  CO  JP  KR  PL  ZA",
      "recall_number": "Z-2818-2020",
      "product_description": "Single RIKKI Retractor, Catalog #71934609 - Product Usage: is used during surgery to hold soft tissue away from the surgical site.",
      "product_quantity": "111 units",
      "reason_for_recall": "The products were improperly heat treated.",
      "recall_initiation_date": "20200803",
      "center_classification_date": "20200817",
      "report_date": "20200826",
      "code_info": "Lot Numbers: 18DM14901, 18FM21215, 18DM07309, 18GM20885, 19EM21519, 18JM05681, 19EM21520"
    }
  ]
}