  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Durham",
      "address_1": "100 Rodolphe St",
      "reason_for_recall": "Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.",
      "address_2": "",
      "product_quantity": "13,961,320 US",
      "code_info": "VITEK 2 AST-P631, Ref # 414961 UDI  03573026426804 OXSF, multiple Gram Positive Cards all lots of cards with Oxacillin and Cefotoxin        OX, OXSF, multiple Gram Positive Cards    all lots of cards with Oxacillin and Cefotoxin    VITEK 2 AST-GP67 Test Kit, Ref: # 22226  UDI  03573026242060, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.  VITEK 2 AST-GP71 Test Kit, Ref: #410750  UDI 03573026355173, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.  VITEK 2 AST-GP75 Test Kit, Ref: #415670  UDI 03573026437183, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.  VITEK 2 AST-GP78 Test Kit, Ref: #421051  UDI 03573026553982, OX, OXSF, multiple Gram Positive Cards, all lots of cards with Oxacillin and Cefotoxin.",
      "center_classification_date": "20180817",
      "distribution_pattern": "Worldwide distribution.  US nationwide including Puerto Rico, Bermuda, Brazil, Canada, China, Curacao, Costa Rica, Dominican Republic, Ecuador, France, Guatemala, Indonesia, Israel, Hong Kong, South Korea, South Africa, Mexico, Myanmar, Nicaragua, Peru, Philippines, Paraguay, Singapore, Thailand, El Salvador, and Vietnam.",
      "state": "NC",
      "product_description": "bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card      Cefotoxin and AST for Oxacillin      bioMerieux VITEK 2 AST-P631  bioMerieux VITEK 2 AST-GP67 Test Kit  bioMerieux VITEK 2 AST-GP71 Test Kit  bioMerieux VITEK 2 AST-GP75 Test Kit  bioMerieux VITEK 2 AST-GP78 Test Kit    Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.",
      "report_date": "20180829",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "bioMerieux, Inc.",
      "recall_number": "Z-2817-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80619",
      "termination_date": "20190520",
      "more_code_info": "",
      "recall_initiation_date": "20180126",
      "postal_code": "27712-9402",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}