      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86104",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2816-2020",
      "product_description": "DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381",
      "product_quantity": "191 units",
      "reason_for_recall": "May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw",
      "recall_initiation_date": "20200721",
      "center_classification_date": "20200817",
      "termination_date": "20220218",
      "report_date": "20200826",
      "code_info": "Lot Number: 29P6036  UDI: 10886982166913"
    }
  ]
}