  "results": [
    {
      "status": "Terminated",
      "city": "Ejido Maneadero",
      "state": "",
      "country": "Mexico",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80460",
      "recalling_firm": "ICU Medical de Mexico, S.A. de C.V.",
      "address_1": "Avenida Cuarzo 250",
      "address_2": "Ensenada",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to AZ, MA, NC, OH, TX, VA, WA.      Worldwide distribution to Belgium, France, Germany.",
      "recall_number": "Z-2816-2018",
      "product_description": "6\" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer.",
      "product_quantity": "300 pouches",
      "reason_for_recall": "ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.",
      "recall_initiation_date": "20180301",
      "center_classification_date": "20180816",
      "termination_date": "20240126",
      "report_date": "20180822",
      "code_info": "Item Number AM6130.      a. Lot Number 3414534, UDI Number (01) 1 0887709 08771 9 (17) 220201 (30) 50 (10) 3414534,  Expiration Date 02/01/2022;    b. Lot Number 3442461, UDI Number (01) 1 0887709 08771 9 (17) 220501 (30) 50 (10) 3442461, Expiration Date 05/01/2022;    b. Lot Number 3509517, UDI Number (01) 1 0887709 08771 9 (17) 220801 (30) 50 (10) 3509517, Expiration Date 08/01/2022",
      "more_code_info": ""
    }
  ]
}