  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72139",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide and internationally, including Canada.",
      "recall_number": "Z-2811-2015",
      "product_description": "Synthes Cranial Flap Tube Clamp and Crimping Device     Product Usage:  The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.",
      "product_quantity": "105882",
      "reason_for_recall": "The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.",
      "recall_initiation_date": "20150831",
      "center_classification_date": "20150923",
      "termination_date": "20160818",
      "report_date": "20150930",
      "code_info": "Part Number  329.315  460.001  460.001.01S  460.002  460.002.01S  460.003  460.003.01S  460.008  460.008.01S  460.009  460.009.01S  460.010  460.010.01S  460.100  460.100.01S  460.107  460.107.01S  with Lot Numbers  4146716, 4154144, 4206682, 4206683, 4213557, 4214824, 4216603, 4216604, 4240689, 4240690, 4241741, 4248462, 4249395, 4249396, 4264386, 4278302, 4278304, 4283035, 4283036, 4283037, 4283038, 4283039, 4283040, 4283041, 4285956, 4289103, 4295084, 4295085, 4298044, 4303093, 4326975, 4326979, 4326981, 4329355, 4331263, 4331265, 4331266, 4344659, 4344678, 4378973, 4482201, 4485317, 4488093, 4488094, 4490608, 4490609, 4496716, 4496915, 4496916, 4501971, 4504043, 4508068, 4513458, 4513459, 4513460, 4513461, 4516371, 4519406, 4519407, 4521408, 4525905, 4550062, 4550063, 4550064, 4550065, 4551164, 4551165, 4551166, 4561400, 4561401, 4572176, 4574481, 4580884, 4586704, 4605094, 4661414, 4669482, 4669483, 4675725, 4695343, 4705970, 4723148, 4753792, 4753794, 4777999, 4778000, 4906228, 4906229, 4914498, 4917681, 4934562, 4944809, 4956361, 4956363, 4983255, 4983256, 4983257, 5007806, 5041856, 5041857, 5041858, 5046857, 5046859, 5098159, 5113496, 5113497, 5134868, 5134869, 5143883, 5204105, 5230669, 5234395, 5234396, 5245233, 5245237, 5251710, 5292433, 5308690, 5308733, 5341509, 5370371, 5370375, 5438364, 5452452, 5465222, 5655435, 5667092, 5697537, 5726966, 5726967, 5726968, 5726969, 5726970, 5726971, 5801281, 5801284, 5801285, 5801287, 6036379, 6036381, 6168066, 6188888, 6188890, 6268389, 6268390, 6272370, 6322541, 6364587, 6553805, 6861415, 6983782, 7032568, 7505227, 7640627"
    }
  ]
}