  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74009",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "1000 Lee Road",
      "address_2": "N/A",
      "postal_code": "14606",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of:   AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.",
      "recall_number": "Z-2804-2016",
      "product_description": "The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876,  is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.",
      "product_quantity": "520 Units",
      "reason_for_recall": "There is an incorrect value (data/calibration mathematics) on ADDs.  This incorrect value will prevent a successful calibration of the assay.",
      "recall_initiation_date": "20160419",
      "center_classification_date": "20160914",
      "termination_date": "20180608",
      "report_date": "20160921",
      "code_info": "Lot 3155  Product Codes: 6842906 & 6801876",
      "more_code_info": ""
    }
  ]
}