  "results": [
    {
      "status": "Terminated",
      "city": "Buffalo Grove",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71928",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "1700 Leider Ln",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Colombia, Czech Republic , Germany, Spain, France, England, Georgia, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lithuania, Myanmar, Mexico, Malaysia, Netherland, Norway, Peru, Philippines, Poland,  Qatar, Reunion, Romania, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Turkey, Taiwan, and South Africa.",
      "recall_number": "Z-2804-2015",
      "product_description": "The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments.    The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.",
      "product_quantity": "U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems.",
      "reason_for_recall": "The ASP6025 Tissue Processor instrument is incorrectly getting into a state of \u001cbottle empty\u001d  during retort filling, after passing the fill lever sensor 2 and prior to reaching level  sensor 3, thus not completing a 5 litre fill.",
      "recall_initiation_date": "20150727",
      "center_classification_date": "20150922",
      "termination_date": "20170105",
      "report_date": "20150930",
      "code_info": "Product Code: Leica ASP6025     Serial Numbers:  163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743,  745, 747, 749, 751, 753, 755, 757, 759, 761, 763, 765,  767, 769, 771, 773, 775, 777, 779, 781, 783, 785, 787,  789, 791, 793, 795, 797, 799, 801, 803, 805, 807, 809,  811."
    }
  ]
}