      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71842",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and to the countries of :  ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, HUNGRY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, MEXICO, NORWAY, POLAND, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN and UNITED KINGDOM.",
      "recall_number": "Z-2801-2015",
      "product_description": "Lightspeed  16 CT",
      "product_quantity": "416",
      "reason_for_recall": "Potential for expulsion of the Power Supply on Lightspeed CT scanner or Discovery ST PET CT or Discovery LS PET CT scanner.",
      "recall_initiation_date": "20150717",
      "center_classification_date": "20150921",
      "termination_date": "20160308",
      "report_date": "20150930",
      "code_info": "00000058433HM4 4860LS16GS  FMI25463004 910213080  FMI25463005 910212101  FMI25463006 910212105  FMI25463007 910213064  FMI25463008 910216041  FMI25463009 910216036  00000321402CN0 910212106  00000325191CN5 AH5362CT02  FMI25463010 190020CT03  FMI25463011 110807CT01  00000980106YM9 110040CT03  00000307018CN2 416249LS16  00000298660CN2 204787LSPD2  00000318927CN1 514468BLS  00000329994CN8 819561LS  00000389961CN4 709891CT  00000335314CN1 514934CLS16  00000340630CN3 82421140148  00000331165CN1 82421010103  00000336002CN1 82421130067  00000334898CN4 82421140150  00000322331CN0 82421160145  00000999187YM8 82421100291  FMI25463133 828210047  FMI25463136 82421310073  00000989604YM4 82421030247  FMI25463015 82421010063  00000075153HM7 82421120171  FMI25463016 082421ATCLSP04  FMI25463017 82421100443  00000080677HM8 82421090111  00000092789HM7 82421120191  00000092771HM5 82421120188  00000074379HM9 82421060038  FMI25463018 82421130055  00000092530HM5 82421110048  00000080298HM3 82421100270  FMI25463019 395291CT01  FMI25463020 PM1633CT04  FMI25463021 PC7479CT02  FMI25463022 DK1025CT02  00000420281CN8 NP384114  FMI25463023 FI2000MN01  00000953833YM1 FI1052CT02  FMI25463024 C6010809  FMI25463025 M1076839  FMI25463026 M40345153  FMI25463027 M4496936  FMI25463028 M4409416  FMI25463029 M2147233  FMI25463030 M4505261  FMI25463031 M4020535  FMI25463032 M4491526  FMI25463033 M2802781  FMI25463034 X971285001  FMI25463035 A4463317  FMI25463036 B4195716  00000320672CN9 M4160064  00000271501CN9 M4182817  00000302750CN5 HC3314CT02  00000311056CN6 PC1268CT03  00000980113YM5 PROS01CT04  00000309302CN8 HC3428CT02  00001033124YM7 HC1560CT05  00000031399CN5 CC0618CT03  00000301205CN1 HC1627CT04  00000000716YC1 PX2261CT02  FMI25463041 GOLD010271  FMI25463042 GOLD010160  00000297465CN7 HU1039CT03  00000313648CN8 HU1008CT02  FMI25463043 835210046  FMI25463045 83021842100841  00000475152YM5 83097598280000  FMI25463046 83021002101343  FMI25463047 83021002101503  FMI25463048 83037594099999  00000342119CN5 83021042101703  00000319062CN6 83021200019012  00000326255CN7 83021002100903  FMI25463134 IL1041CT01  FMI25463044 IL1000CT01  FMI25463049 A5177356  00000974327YM9 A5415714  FMI25463050 A5375119  00000297467CN3 A5315535  00000298788CN1 A5404503  00000298327CN8 A103421302  00000317806CN8 A5152959  00000298364CN1 A5109820  00000317976CN9 QW3044  00000277446CN1 QY0024  00000941020YM0 QW3077  00000311054CN1 QW3036  00000953849YM7 QWZ057  00000941010YM1 XF0018  00000953829YM9 QT3033  00000980097YM0 QW3059  00000305242CN0 QW2025  00000988399YM2 QW3064  00000297463CN2 QY0006  00000953847YM1 QT3036  00000304158CN9 QY0002  00000310122CN7 QY0058  00000980099YM6 QZ0001  00000941033YM3 QT3032  00000303514CN4 QW3033  00000988404YM0 XF0149  00000994079YM2 QT4051  00000980096YM2 QT4049  00000941031YM7 QT3031  00000328077CN3 QW3045  425784CN6 HD0041  33373GM2 RY0028  239249HM6 BD9217  FMI25463051 QT4074  FMI25463052 QT3042  FMI25463053 XF0022  FMI25463054 QW3049  00000296397CN3 QW2016  FMI25463055 QZ0120  00000004875YC1 QZ0066  FMI25463056 YN0382  FMI25463057 QW3057  00000297397CN2 QW3043  00001023672YM7 QW3073  00000314641CN2 850210549  00000067961HM3 850210557  00000071678HM7 850210558  00000086904HM0 850210826  00000066382HM3 850210555  00000392647CN4  LSPRO2643765  00000295414CN7 73801LS16S  00000307947CN2 NO1052CT02  00000286517CN8 604853CTLS  FMI25463135 600042CT01  FMI25463058 809780EXCITE  FMI25463059 SA1279CT01  00000296276CN9 SA1011CT02  FMI25463060 Z001CT02  1380806 847210020  00000303589CN6 SK1003CT01  00000310270CN4 ZA1247CT01  00000994088YM3 ZA1127CT02  00000318926CN3 ZA2471CT01  00000205583HM8 ZA1058CT02  00000309308CN5 DEN27243  00000030411CN8 DPM32132  00000329991CN4 DEL89957  00000329306CN5 DE188430  00000310874CN3 MPX02003  00000327272CN1 DEL26410  00000297385CN7 DPB15832  00000304109CN2 FPG08400  00000297902CN9 FPG03601  00000279527CN6 FPG81004  00000299766CN6 A010CT07  00001033115YM5 547110CT02  FMI25463012 475230CT04  FMI25463013 120928CT01  FMI25463014 115030CT01  00000312924CN4 00160CTS02  00000336000CN5 10649CTS01  00000294856CN0 00361CTS02  00000984693YM2 00664CTS02  00000302754CN7 10628CTS01  313653CN8 70328CTS03  FMI25463037 00537CTS01  FMI25463038 80040CTS02  FMI25463039 00314CTS02  FMI25463040 10470CTS01  309307CN7 SHICT18  00000312257CN9 503215PCT4  00000293010CN5 503DRCCT  00000296272CN8 518828CT  00000323860CN7 ALLIANCECT74  00000318314CN2 949949CT  00000310267CN0 715453CT1  00000317790CN4 612863LSC  00000321966CN4 732671LS16  00000338403CN9 850718JLS  00000336255CN5 702880CT  00000339629CN8 914681PRO  00000334464CN5 RADNETCT1483  00000297390CN7 518438LS16  00000362635CN5 615320CT  00000341886CN0 305285LS  00000314491CN2 252847CT3  00000321140CN6 773792PRO  00000331769CN0 775785CT16  00000320666CN1 972548LS16  00000313370CN9 434243UVICT1  00000311424CN6 770751QXI  00000297382CN4 206232CT1  00000299234CN5 RADNETCT1500  00000292531CN1 423272CT  00000296274CN4 574371CT16  00000288654CN7 703385LS16  00000277800CN9 415723SCT3  00000348034CN0 410337LS16  00000318730CN9 480237CT16  00000328358CN7 4147614XCT1  00000338570CN5 304526LSPRO  00000308842CN4 813870LS16  00000307026CN5 847437ACT4  00000332709CN5 940384MDCT  00000347365CN9 770736LCT  00000314952CN3 718918LS16  00000293011CN3 CT81441  00000313365CN9 706845WGLS  00000308983CN6 918494SFLS2  00000297426CN9 412777LS16  00000307024CN0 956969CT  00000324976CN0 770778CTP16  00000318876CN0 918683LS16  00000338401CN3 617355CT3  00000294503CN8 770718OCT1  00000347517CN5 509248VCT16  00000319953CN6 757886CT  00000351700CN0 RADNETCT1404  00000294766CN1 402481BWCT1  00000323330CN1 804864BRLS16  00000315054CN7 907729CTPRO  00000293499CN0 212860LS16  00000299763CN3 208467CTLS  00000299311CN1 910215MR16  00000320667CN9 281PERWCT  00000313649CN6 423586CT2  00000294504CN6 949760CT2  00000321162CN0 760775CT  00000298790CN7 ALLIANCECT67  00000300371CN2 573635SMCT2  00000299765CN8 703578LS16  00000331098CN4 425261CT  00000342628CN5 773967LS16  00000304113CN4 276666CT  00000293037CN8 631475LS16  00000295005CN3 785452SCT  00000296280CN1 805SVCT1  00000336852CN9 772335CT1  00000362634CN8 570522CT1  00000351636CN6 845334LS  00000316660CN0 402887CT16  00000308839CN0 309672MCT1B  00000349343CN4 281395LS16  00000320654CN7 352728FHVC1  00000295410CN5 919731CT16  00000333823CN3 330830LS16J  00000332736CN8 206731ERCT  00000308361CN5 972420LS16  00000329988CN0 912692LS3  00000346883CN2 281FCLPCT  00000349345CN9 605342RCMC  00000298345CN0 734240CT16  00000323050CN5 781340SCT4  00000332081CN9 401769WLS  00000310731CN5 904391CTLS  00000326112CN0 406293LCT  00000317788CN8 ALLIANCECT127  00000311417CN0 281274CT3  00000366432CN3 253944CT16  00000342587CN3 207777LS16  00000320668CN7 724537CTLS  00000301084CN0 914737CTLS  00000294344CN7 504873TL16  00000350494CN1 303695LS16  00000298360CN9 210292LSA  00000314950CN7 913541CT16  00000338127CN4 801773CT2  00000288959CN0 812471CT  00000285030CN3 843236CFCT  00000290655CN0 937534CT  00000843255YM1 916RASCCT  00000886944YM8 336433KV16  00000327821CN5 918294SCLS  FMI25463001 412469CTH1  00000358633CN6 815338PRO  00000886897YM8 912355LS16  00000273706CN2 423698DLS  00000281948CN0 SHICT288  00000361749CN5 503513PCT16  00000289411CN1 812537LS16  00000279106CN9 302653LS16  00000267340CN8 513856CT  00000279054CN1 609652LS16  00000271998CN7 817551OPCT  00000176078HM4 989839RT16  FMI25463002 610270LSPRO  00000265944CN9 940328LS16  00000296243CN9 925813CTM  00000919623YM9 912748CT1  00000265771CN6 918224LSCT  00000278570CN7 478745CT  00000328341CN3 863687LS8ER  00000385174CN8 814624LS1  00000367539CN4 432685LS1  00000843223YM9 RADNETCT1503  FMI25463003 860444CTPRO  00000381259CN1 303340CTHP  00000285331CN5 850747GCCT  00000276921CN4 704362MQCT  00000376596CN3 775885PRO  00000269963CN5 903TSJCT16  00000379714CN9 208455CTLS  00000295008CN7 214345LS16  00000984695YM7 515247CT5  00000386311CN5 623207CT16  00000988390YM1 864225CT  00000984689YM0 9167345X  953828YM1 SHICT19  00000953839YM8 305592LS16  00000358180CN8 505848LCT  00000313368CN3 609584LS16  00000387780CN0 631MSKLS16  00000314494CN6 713566LCT2  00000350423CN0 904824FCT4  00000365567CN7 208524PETCT  00000382184CN0 252641HCT16  00000328883CN4 317745LSPRO  00001030388YM1 816691PCT2  00000360963CN3 858657RTCT  00000984696YM5 901726CT2  00000365453CN0 907459TVCT  00000317093CN3 940384LS16  00000766980YM7 205601BHCT  00000263412HM9 716UROCT  00000940961YM6 845887LS16  00000412581CN1 414541VCT1  00000300854CN7 605882CT  00000308734CN3 00000074379HM9  00000315227CN9 607734H16  00000313363CN4 765423LS16  00000308836CN6 214345LINUX16  00000330532CN3 561650CT1  00000298793CN1 847LS16CT1  00000882908YM7 919CHTMCT  00000345928CN6 718920LSRO  FMI25463061 760242LS16  FMI25463062 DCOSDI200000094  FMI25463063 423224CT2  FMI25463064 435251CT1  FMI25463065 281258PERMB  FMI25463066 DCOSDI400000077  FMI25463067 ALLIANCECT103  FMI25463068 ALLIANCEPCT85A  FMI25463069 580323LSP  FMI25463070 510889CT  FMI25463071 215481LS16  FMI25463072 716839LS16  FMI25463073 617726PR12  FMI25463074 662538CT  FMI25463075 916734CT1  FMI25463076 425261CT2  FMI25463077 DCOSDI100000086  FMI25463078 630325DLS  FMI25463079 516520LS16  FMI25463080 812450PRO100  FMI25463081 SHAREDPET2  FMI25463082 717273CT1  FMI25463083 301496CTPET  FMI25463084 972579L16A  FMI25463085 608364ERCT  FMI25463086 605622CT1  FMI25463087 716898CT  FMI25463088 6036503T  FMI25463089 720FCACT1  FMI25463090 662756CT  FMI25463091 915EPDST1  FMI25463092 918786LS1  FMI25463093 435734CT2  FMI25463094 212639LS03  FMI25463095 706316CT1  FMI25463096 DCOSDI100000423  FMI25463097 765646LS16  FMI25463098 201330LS16  FMI25463099 715387CT2  FMI25463100 404329LS16  FMI25463101 DCOSDI200000112  FMI25463102 916784CT1  FMI25463103 DCOSDI000000343  FMI25463104 717633CT2  FMI25463105 313916LS16  FMI25463106 317582LSPRO  FMI25463107 SMRDST6  FMI25463108 304255LS16  FMI25463109 DCOSDI300000210  FMI25463110 3411  FMI25463112 901729MSCT  FMI25463113 209536CT1  FMI25463114 ALLIANCEPCT71P  FMI25463115 303398CTLS16  FMI25463116 916734CT2  FMI25463117 724832CT  FMI25463118 MISECT33L  FMI25463119 734936CT1  FMI25463120 3155  FMI25463121 915670HLS2  FMI25463122 205759CTI2  FMI25463123 308537GH16  FMI25463124 870972LS16  FMI25463125 276679CTLS  FMI25463126 317846LS  FMI25463127 952924CT1  FMI25463128 206731CT1  FMI25463129 956350VRCT  FMI25463130 847ARLS  FMI25463131 610954CT16  FMI25463132 608263CT3  FMI25463137 919684B5A  FMI25463138 919684C3A  FMI25463139 319272LS16  FMI25463140 408757LS16  FMI25463141 630275LSER  FMI25463142 760830CT  FMI25463111 DCOSDI000000013"
    }
  ]
}