  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72038",
      "recalling_firm": "Zien Medical Technologies, Inc",
      "address_1": "2500 S State St",
      "address_2": "Sted240",
      "postal_code": "84115-3110",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution to U.S. nationwide, Germany and England.",
      "recall_number": "Z-2800-2015",
      "product_description": "PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15",
      "product_quantity": "28,284",
      "reason_for_recall": "Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.",
      "recall_initiation_date": "20150818",
      "center_classification_date": "20150921",
      "termination_date": "20151209",
      "report_date": "20150930",
      "code_info": "Product Code PB-M-10, Lots 140901J-A, 150228PB10, 150509PB10, 150510PB10; Product Code PB-M-11P, Lots 140901J-B, 150227PB11, 150318PB11, 150509PB11, 150510PB11; Product Code PB-M-15, Lots 140901J-C, 150226PB15, 150326PB15, 150511PB15, 150512PB15, 150529PB15."
    }
  ]
}