      "status": "Terminated",
      "city": "Shanghai",
      "state": "N/A",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85948",
      "recalling_firm": "Shanghai United Imaging Healthcare Co., Ltd.",
      "address_1": "No. 2258 Chengbei Rd, Jiading Ind.",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution including in the state of Texas.",
      "recall_number": "Z-2799-2020",
      "product_description": "uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.",
      "product_quantity": "1 unit distributed to Texas",
      "reason_for_recall": "1)\t1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.",
      "recall_initiation_date": "20200716",
      "center_classification_date": "20200813",
      "termination_date": "20220707",
      "report_date": "20200819",
      "code_info": "Serial No. F00002; GTIN 06971576832040"
    }
  ]
}