  "results": [
    {
      "status": "Terminated",
      "city": "Bonaduz",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80467",
      "recalling_firm": "Hamilton Medical AG",
      "address_1": "Via Crusch 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI.      Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.",
      "recall_number": "Z-2796-2018",
      "product_description": "Hamilton-C3 Ventilator Unit    Product Usage:  The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its configuration. The device is used in the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room, and during transfer of ventilated patients within the hospital.",
      "product_quantity": "2 devices",
      "reason_for_recall": "An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.",
      "recall_initiation_date": "20180309",
      "center_classification_date": "20180813",
      "termination_date": "20190318",
      "report_date": "20180822",
      "code_info": "Serial Number: 3084, 3088",
      "more_code_info": ""
    }
  ]
}