  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75015",
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide and Internationally",
      "recall_number": "Z-2789-2016",
      "product_description": "Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10)    The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.",
      "product_quantity": "2,039 units",
      "reason_for_recall": "Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.",
      "recall_initiation_date": "20160801",
      "center_classification_date": "20160913",
      "termination_date": "20170608",
      "report_date": "20160921",
      "code_info": "Catalog No. 3182-1000  Lot #'s: RD8E100, RD8A054TT, RD8A054T, RD8E099M, RD8E099A, RD8E099, RD8A054A, RD8E099D, RD8A053A, RD8A054, RD8A053T, RD7K062L, RD8A052A, RD8A052E, RD8A052, RD8A053, RD7E128D, RD5T290H, RD7E128T, RD5S262D, RD5T290W, RD7K062E, RD5S262X, RD7E128Y, RD7A076MA, RD6W102MA, RD7K062A, RD7K062, RD7E128H, RD7A076A, RD7E128A, RS7E128, RD6W103D, RD6V108J, RD6V108K, RD6V108H, RD7A076H, RD6V112M, RD7A074D, RD7A076J, RD6V112H, RD6W102m, RD5H286A, RD2V022, RD2K049, RD6W103, RD5H286J, RD2T037A, RD7A076M, RD6S136A, RD5H286, RD2V031C, RD5N361T, RD6V112A, RD4M068J, RD2V021C, RD7A076, RD6V112, RD4E226W, RD2V021, RD7A074H, RD6W102, RD4S220W, RD2V023X2, RD6W103A, RD6S136M, RD4M069D, RD2V023X3, RD6V108, RD5W148M, RD4S220, RD2V023, RD7A074, RD5M237K, RD4S220A, RD2V031A, RD7A074A, RD5T290X, RD4S220A1, RD2V023X1, RD6W103M, RD5T290A, RD4M069, RD2V031, RD5K116A, RD6S137A, RD4L128X1, RD2V031X2, RD5M236T, RD5W148A, RD4L128V, RD2V031X1, RD5K116T, RD6S137, RD4E226J, RD2N030K, RD3V002A, RD6S136, RD4M068< RD2T037, RD5M237, RD5W148X, RD4L128, RD2T037X, RD5K116N, RD5W148, RD4E226M, RD2N030E, RD5K116, RD5T290M, RD4E226L1, RD2N030S, RD5M236, RD5T290E, RD4C230L, RD2N030X, RD5H286L, RD5T290, RD4E226L, RD2N030XX, RD3T155L, RD5T290D, RD4E226, RD2N030, RD2V021E, RD5N361A, RD4C162L, RD2K068W, RD2V022W, RD5S305J, RD4C162, RD2N030X1, RD2V021S, RD5S305M, RD4C230, RD2M019E, RD2V022K, RD5N361M, RD4C230X, RD2M039, RD2V022E, RD5S305, RD3V002P, RD2M019H, RD2V021H, RD5S262A, RD3V002X1, RD2K049H, RD2V023E, RD5S262M1, RD3V002X2, RD2M003H, RD2T037W, RD5S262M, RD3V002D1, RD2M038H, RD2K068X1, RD5S262J, RD3V002, RD2K068B, RD2K068, RD5S262, RD3V002D, RD2M003, RD2K049R, RD5N361H, RD3V206D, RD2M038, RD2K049E, RD5M259M, RD3T160L1, RD2M019, RD2K017E, RD5N316, RD3V206X1, RD2K068S, RD2H056, RD5H286D, RD3T155L1, RD2K056S, RD2K056, RD5M259T, RD3T160L3, RD2K065S, RD2K065, RD5M236H, RD3T160, RD2K049S, RD2K017, RD5M259, RD3T160L2, RD2K068E, RD5M237L, RD3T160L, RD2H056E, RD5K116J, RD3V206, RD2K065E, RD4S220D, RD3T155 and RD2K056E",
      "more_code_info": ""
    }
  ]
}