  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71477",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to state of WA; and countries of: France, Sweden, Austria, Spain, Germany, Italy, UK, New Zealand.",
      "recall_number": "Z-2785-2015",
      "product_description": "Radiation Therapy Treatment Planning System software    RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.",
      "product_quantity": "20 devices",
      "reason_for_recall": "An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables  for dose, dose difference and PET images",
      "recall_initiation_date": "20150529",
      "center_classification_date": "20150917",
      "termination_date": "20170313",
      "report_date": "20150923",
      "code_info": "software build numbers are 4.7.0.15 or 4.7.1.10",
      "more_code_info": ""
    }
  ]
}