{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74835",
      "recalling_firm": "The Anspach Effort, Inc.",
      "address_1": "4500 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and including the countries of Canada, Austria, Australia, Belgium, Brazil,  Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Libya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, Slovakia, Spain, South Africa, South Korea, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam, and areas of Middle East, Africa, Asia & Pacific.",
      "recall_number": "Z-2783-2016",
      "product_description": "Light Adaptor for Small Battery Drive and Small Battery Drive II.",
      "product_quantity": "1,833 devices (05.001.024 and 05.001.108)",
      "reason_for_recall": "There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.",
      "recall_initiation_date": "20160624",
      "center_classification_date": "20160926",
      "termination_date": "20180609",
      "report_date": "20161005",
      "code_info": "Model/Catalog:  Part # (05.001.108 and 05.001.024)  Serial:  All Serial Numbers."
    }
  ]
}