      "country": "United States",
      "city": "Buffalo Grove",
      "address_1": "1700 Leider Ln",
      "reason_for_recall": "The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.",
      "address_2": "",
      "product_quantity": "901",
      "code_info": "Product Code: Leica ASP200S/ASP300S with Serial Numbers:  5682; 5886; 5264; 5267; 5356; 5259; 5956; 5946;  5794; 5385; 5929; 5301; 5401; 5310; 5530; 5909;  5804; 5272; 5724; 5796; 5807; 5446; 5435; 5670;  5947; 5348; 5434; 5388; 5366; 6000; 6131; 5407;  6040; 5799; 5815; 5737; 5706; 5381; 5416; 5673;  5875; 5962; 6041; 6042; 5266; 5948; 5693; 6021;  5610; 5798; 5926; 5387; 5349; 5273; 6115; 5612;  5965; 5955; 5260; 5725; 5871; 5873; 5874; 5562;  6046; 5386; 5977 ;5889; 5398; 6062; 5609; 6015;  5734; 5403; 5289; 5426; 6149; 5663; 6045",
      "center_classification_date": "20150917",
      "distribution_pattern": "Worldwide Distribution - US including U.S. including AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, TX, VA, WA,  WI, WV and District of Columbia. Internationall to AE, AT, AU , BD , BE, BF, BG,BR, CA, CL, CN, CO, CR, DE, DK, EC, EG, ES, ET, FI, FR, GB, HU, ID, IN, IT, JM, JP, KR, LK, MA, MT, MY, NL, NZ, PA, PE, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, TW, VN and ZA.",
      "state": "IL",
      "product_description": "The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments.  The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.",
      "report_date": "20150923",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Leica Microsystems, Inc.",
      "recall_number": "Z-2779-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71910",
      "termination_date": "20170105",
      "more_code_info": "",
      "recall_initiation_date": "20150805",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}