  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dganya Bet",
      "state": "N/A",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80641",
      "recalling_firm": "Degania Silicone, Ltd.",
      "address_1": "Degania Bet",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.",
      "recall_number": "Z-2778-2018",
      "product_description": "Thermistor Foley catheter, nonsterile, Catalogue Number 102201101663MD    Product Usage:  This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.",
      "product_quantity": "14326",
      "reason_for_recall": "The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.",
      "recall_initiation_date": "20180621",
      "center_classification_date": "20180920",
      "termination_date": "20241025",
      "report_date": "20180926",
      "code_info": "V17012299, V17012518, V17012896, V17013267, V17013268, V17013272, V17023765, V17025022, V17025023, V17026861"
    }
  ]
}