  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74940",
      "recalling_firm": "Alere San Diego, Inc.",
      "address_1": "9975 Summers Ridge Rd",
      "address_2": "N/A",
      "postal_code": "92121-2997",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution - U.S Nationwide in the states of:  TN, MA, GA, OK, VA, FL, TN, MO, AL, TX, IL, NY, VT, RI, KY, WI, AL, NC, LA, SD, TX, VA, MI, MN, IN, IA, CA, AR, PA, HI, WV, KS, CO, CT, UT and the countries of Japan, England, Canada",
      "recall_number": "Z-2778-2016",
      "product_description": "LDX Optical Check Cassettes (OCC)  Model Number: 10-228 and 14-535    Cassette in box with foam insert. The cassette is used to check the optics system of the LDX analyzer.    Product Usage:  The Optics Check Cassette is used to check the optics system of the Alere Cholestech LDX¿ Analyzer. It is run once a day before patient samples are tested and after the analyzer has been moved or serviced.",
      "product_quantity": "12,951",
      "reason_for_recall": "Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.",
      "recall_initiation_date": "20160802",
      "center_classification_date": "20160909",
      "termination_date": "20161028",
      "report_date": "20160921",
      "code_info": "Model Number: 10-228 and 14-535  Lot Number: 332994",
      "more_code_info": ""
    }
  ]
}