  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lewistown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77756",
      "recalling_firm": "GE Inspection Technologies, LP",
      "address_1": "50 Industrial Park Rd",
      "address_2": "N/A",
      "postal_code": "17044-9312",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : AL, SC, OH, MI, PA, and Internationally to Germany and Singapore",
      "recall_number": "Z-2777-2017",
      "product_description": "GE Inspection Technologies Blade|line Industrial Cabinet X-ray system",
      "product_quantity": "13 units distributed in the US",
      "reason_for_recall": "GE Inspection Technologies, LP determined that the Blade|line cabinet x-ray system fails to comply with the cabinet x-ray performance standard\u0019s emission limit of 0.5 milliRoentgen in one hour (21 CFR ¿ 1020.40(c)(1)(i)) underneath the bottom center of the system.  The crawl space affected by the excess emission is approximately 22 inches high, normally covered by decorative panels, and is not intended to be accessed.",
      "recall_initiation_date": "20170628",
      "center_classification_date": "20170801",
      "termination_date": "20200206",
      "report_date": "20170809",
      "code_info": "Serial No.    TM  201317  TM  201312  TM  201323  TM  201331  TM  201401  TM  201329  TM  201404  TM  201413  TM  201414  TM  201415  TM  201424  TM  201425  158017R1"
    }
  ]
}