  "results": [
    {
      "status": "Terminated",
      "city": "Dganya Bet",
      "state": "N/A",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80641",
      "recalling_firm": "Degania Silicone, Ltd.",
      "address_1": "Degania Bet",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.",
      "recall_number": "Z-2774-2018",
      "product_description": "Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101880MD    Product Usage:  This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.",
      "product_quantity": "500",
      "reason_for_recall": "The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.",
      "recall_initiation_date": "20180621",
      "center_classification_date": "20180920",
      "termination_date": "20241025",
      "report_date": "20180926",
      "code_info": "S17003321"
    }
  ]
}