      "country": "United States",
      "city": "Irvine",
      "address_1": "33 Technology Dr",
      "reason_for_recall": "Advanced Sterilization Products (ASP) is reminding all CIDEX OPA Solution customers to follow the warnings and rinsing procedure in their current labelling around the criticality of proper rinsing of medical device following disinfection with CIDEX OPA according to the CIDEX OPA Solution IFU (Instruction For Use).",
      "address_2": "",
      "product_quantity": "318,632 cases",
      "code_info": "010713325  010713325  010713324  010713326  010713324  020713327  010713327  010713327  020713327  090713339  020713331  090713339  090713343  090713343  090713345  100713347  110713352  120713353  120713352  100713347  110713352  120713353  150713355  150713356  150713354  160713360  150713356  150713354  150713354  170713361  170713363  170713362  170713363  170713361  170713364  180713368  190713368  190713369  180713368  190713369  290713370  290713370  050813376  050813377  050813377  050813378  070813384  060813379  050813378  070813384  070813383  070813383  070813385  120813391  120813393  070813385  120813392  120813393  140813399  140813400  140813398  140813399  130813398  190813406  190813408  120813391  130813398  120813394  190813406  200813412  210813414  210813415  210813414  210813413  210813415  230813421  230813422  230813422  270813424  270813420  270813423  270813424  290813430  270813420  140813400  190813407  190813408  290813429  290813432  290813429  290813430  290813431  300813435  290813432  290813433  300813435  210813413  290813431  020913434  020913434  020913438  020913438  040913444  040913445  020913438  040913444  030913442  040913443  040913445  060913450  060913450  090913451  090913451  090913452  110913458  110913457  110913459  110913457  160913466  160913466  160913467  180913472  180913473  160913464  160913467  180913472  180913471  180913473  180913474  230913478  180913474  230913478  230913479  230913479  230913480  240913482  240913482  250913485  260913489  260913488  250913485  250913486  260913488  300913490  300913491  260913489  300913490  100913456  110913458  090913449  090913452  100913456  021013496  021013497  021013497  071013503  021013495  071013503  071013504  071013501  071013504  071013501  091013509  091013510  021013496  141013511  141013514  141013516  091013510  141013514  161013521  161013522  141013511  161013521  161013520  161013522  211013523  211013523  211013528  211013527  211013528  221013530  221013530  231013535  231013533  231013535  231013533  241013536  281013540  281013541  281013539  281013541  241013536  241013538  241013538  241013537  241013538  281013542  281013539  281013542  311013553  011113553  041113554  041113554  041113556  041113556  041113554  041113557  041113558  081113569  111113570  041113558  081113569  111113568  111113570  111113571  131113576  131113577  111113568  131113577  151113585  181113586  181113587  131113579  151113585  131113578  131113579  201113593  201113594  181113584  181113587  181113584  181113588  201113592  201113594  201113595  201113595  261113603  251113603  251113599  251113603  261113605  261113606  011113555  201113593  011113555  201113593  261113606  271113608  271113611  271113608  021213616  021213617  021213615  021213617  021213615  021213615  051213625  061213627  061213627  061213628  091213631  091213631  091213632  101213637  111213637  111213639  161213648  161213648  151213647  161213645  161213649  111213639  111213640  161213649  161213650  161213650  171213651  171213654  171213654  181213655  181213655  181213656  071213630  091213632  191213661  201213661  181213657  181213658  181213658  191213659  191213659  191213660  191213661  051213625  051213626  111213641  151213647  111213641  111213638  111213640  111213641  060114007  070114004  020114002  020114003  030114003  060114007  070114004  080114012  090114013  020114001  020114001  020114002  090114014  090114015  090114013  090114014  090114015  130114021  130114022  130114022  140114024  140114024  140114027  150114029  150114029  150114030  180114036  150114030  180114036  200114034  200114037  200114034  200114038  200114038  200114039  210114040  060114006  220114045  060114006  210114040  220114045  230114045  270114047  270114048  270114048  270114049  270114049  290114053  290114055  290114055  310114056  290114055  310114056  310114057  310114057  310114058  010214059  020214061  030214061  010214059  030214060  030214061  030214062  030214062  030214063  030214064  030214060  030214063  030214064  040214064  050214070  050214069  050214070  070214077  060214070  070214077  100214078  100214075  100214079  100214075  100214080  120214085  120214087  100214080  120214086  120214087  120214085  120214088  140214096  180214100  120214088  180214100  180214101  180214101  180214102  200214108  200214108  200214109  200214109  210214110  270214123  270214126  270214126  210214111  270214123  270214123  210214110  030314127  030314130  030314129  030314130  030314127  030314131  100314142  100314147  100314145  100314142  100314146  070314144  100314145  070314143  070314144  120314154  130314156  130314155  130314156  130314157  170314158  170314163  140314159  130314155  100314147  120314152  120314153  120314152  120314154  140314159  140314160  170314162  170314158  170314162  170314163  190314169  190314170  190314170  250314182  250314184  260314185  260314187  250314184  260314187  260314185  280314192  280314193  280314192  280314193  310314194  310314194  310314191  310314195  310314191  240314176  250314182  070314143  050314136  060314141  050314136  050314138  030314131  050314137  050314138  070414202  070414202  080414208  090414211  090414212  090414211  090414213  100414215  100414215  150414224  150414224  150414225  150414225  150414226  150414223  150414226  150414227  160414227  160414227  170414233  170414234  170414234  170414235  170414235  180414235  220414236  220414232  220414236  220414237  220414238  230414239  230414242  230414242  240414244  240414244  240414244  240414245  220414232  220414238  240414246  280414247  240414243  240414245  240414246  240414245  240414246  240414243  290414252  290414255  280414247  280414251  290414252  010514262  300414256  300414259  010514260  010514262  290414255  300414256  010514260  010514263  080414206  090414212  080414208  080414206  090414212  4126A  4128A  4132A  060514266  060514267  060514265  060514266  060514265  060514267  120514277  060514268  140514283  190514289  190514291  190514289  190514292  190514292  210514298  210514299  210514297  210514299  210514300  220514303  220514302  220514301  220514303  210514297  210514300  220514301  220514304  230514304  270514307  140514283  140514285  140514283  140514285  140514284  140514285  290514312  290514314  290514313  290514314  300514316  120514276  120514277  140514284  120514278  120514276  270514305  270514307  270514308  270514305  290514313  270514305  040614324  040614325  030614321  040614325  040614323  030614321  030614322  030614320  030614322  040614324  090614326  100614330  100614330  100614331  130614340  160614341  100614329  130614340  100614331  160614342  160614343  160614339  160614341  230614350  230614351  240614354  240614355  240614353  240614355  230614351  230614347  240614353  240614356  260614361  260614361  260614362  260614360  260614362  260614362  260614360  300614365  300614366  300614366  300614363  010714368  300614363  160614343  180614348  180614348  180614349  190614349  230614350  230614352  230614352  020714373  070714381  090714389  100714391  090714389  100714390  100714391  100714392  070714376  070714380  070714379  070714380  100714390  140714395  140714395  140714394  140714396  160714403  160714403  170714405  170714405  170714406  180714408  180714408  180714409  020714373  070714379  230714414  220714412  180714409  230714414  240714418  240714418  240714419  240714419  240714420  040814422  040814422  040814423  040814423  110814428  080814427  110814428  110814429  110814426  110814429  110814426  110814430  180814443  190814442  180814442  180814440  180814443  130814436  130814436  130814437  140814437  180814440  210814450  210814452  190814442  200814447  200814449  220814454  260814454  260814455  220814453  220814454  260814455  260814456  260814457  260814457  270814457  270814459  270814458  270814459  200814447  210814447  210814452  270814458  280814460  290814464  290814464  290814465  290814464  290814465  290814465  030914467  030914468  020914463  030914463  030914467  020914463  020914466  030914468  100914486  110914486  110914488  120914490  120914492  130914490  110914488  120914491  120914491  120914487  120914492  110914487  170914503  170914504  130914490  130914493  150914495  150914494  150914496  150914494  150914495  170914502  170914504  170914505  170914505  180914507  180914508  180914507  150914496  150914497  170914503  180914506  180914508  190914510  190914510  190914511  220914514  230914515  230914515  230914518  230914518  240914520  240914521  190914511  220914509  220914514  240914521  250914525  240914522  250914523  240914519  240914521  260914528  250914524  260914528  250914523  260914528  260914529  260914529  290914531  270914530  290914532  290914527  290914531  300914527  290914532  300914536  110914489  300914536  250914526  110914489  061014544  061014548  031014546  061014545  061014547  061014544  061014547  031014546  061014548  091014556  091014557  091014555  091014557  091014555  091014558  101014560  101014560  101014561  101014559  101014561  101014559  131014564  131014565  141014567  131014565  141014567  141014570  201014582  211014584  211014584  211014585  231014590  231014591  211014585  231014590  211014583  231014589  231014591  301014599  301014600  301014600  311014600  4321A  4322A  4323A  4324A  4325A  031114603  031114602  031114605  031114603  031114602  031114604  031114605  031114606  041114609  041114608  041114609  071114620  101114619  101114623  101114623  101114622  101114619  071114620  101114621  101114622  141114638  171114639  131114633  131114635  141114638  121114629  131114634  131114634  131114635  171114637  171114639  171114640  171114637  171114641  171114641  241114654  241114653  241114654  241114655  241114655  241114653  271114664  011214667  011214669  011214667  011214667  011214668  271114664  011214669  281114666  281114666  4335A  4336A  4337A  4338A  4339A  081214681  081214682  041214677  051214680  081214681  041214677  081214682  111214690  111214691  111214689  111214691  111214689  161214696  161214695  161214694  171214698  161214694  161214696  060115700  060115697  190115018  190115017  060115700  060115701  190115017  060115701  150115016  190115018  220115026  270115028  260115027  270115028  270115029  270115024  270115023  270115024  270115029  280115032  280115031  280115032  290115032  270115023  5033A  5034A  5035A  5036A  5037A  5040A  5049A  5050A  5051A  5054A  030215030  090215047  090215044  100215048  090215044  090215046  030215030  030215035  030215035  020215033  030215034  100215048  110215053  110215052  160215058  110215053  160215058  160215056  230215068  170215061  160215059  180215065  230215070  020215033  170215062  180215064  170215063  170215061  170215062  180215065  180215064  110315093  110315095  170315106  170315108  180315112  170315109  180315110  180315111  180315112  230315121  230315119  250315129  250315130  110315095  110315094  250315129  230315123  110315093  110315096  170315106  170315109  230315119  230315123  230315122  230315121  180315114  180315111  180315113  5117A  5118A  5119A  280415194  280415191  290415198  290415195  290415196  290415195  290415197  300415199  300415201  280415194  290415196  290415198  300415200  300415199  010515203  010515203  5120A  5121A  5124A  060515206  060515204  070515209  060515207  070515209  070515210  060515206  060515205  060515205  050515202  140515215  140515214  140515215  140515212  010615228  010615227  010615225  070515210  140515214  070515208  260515219  260515218  010615227  010615226  260515219  280515216  280515216  010615226  280515221  010615228  050615243  040615241  030615237  030615234  030615235  030615234  030615236  020615230  040615241  030615237  040615238  020615230  030615235  220615275  220615276  220615277  050615243  050615244  100615254  080615246  150615260  150615261  150615262  050615244  060615242  060615245  080615247  060615242  080615246  080615247  080615248  100615254  100615255  100615255  110615257  100615253  150615263  150615260  170615270  180615272  170615271  150615263  170615270  220615278  220615275  150615261  110615257  220615276  180615272",
      "center_classification_date": "20150916",
      "distribution_pattern": "US",
      "state": "CA",
      "product_description": "CIDEX OPA Solution, P/N 20390 for use as a high level disinfectant for reprocessing heat sensitive medical devices.",
      "report_date": "20150923",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Advanced Sterilization Products",
      "recall_number": "Z-2774-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71887",
      "termination_date": "20160520",
      "more_code_info": "",
      "recall_initiation_date": "20150810",
      "postal_code": "92618-2346",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}