  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Canton",
      "address_1": "5449 Research Dr",
      "reason_for_recall": "Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.",
      "address_2": "",
      "product_quantity": "93 kits",
      "code_info": "C7539-150  Lot:  406201  Batch:  279  Expiration date:  2016-03",
      "center_classification_date": "20150916",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.",
      "state": "MI",
      "product_description": "Pointe Scientific Creatinine Reagent  Catalog number C7539-150   in vitro diagnostic  C7539-150 kit includes one R1 and one R2 bottle.  R1:  HDPE plastic bottle. 120 ml fill.  R2:  HDPE plastic bottle.  90 ml fill.",
      "report_date": "20150923",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Pointe Scientific, Inc.",
      "recall_number": "Z-2773-2015",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "71974",
      "termination_date": "20151029",
      "more_code_info": "",
      "recall_initiation_date": "20141212",
      "postal_code": "48188-2261",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}