{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Maple Grove",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99236",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "1 Scimed Pl",
      "address_2": "N/A",
      "postal_code": "55311-1565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.",
      "recall_number": "Z-2771-2026",
      "product_description": "Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.",
      "product_quantity": "68 units",
      "reason_for_recall": "Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects \"Recalculate\",  the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.",
      "recall_initiation_date": "20260619",
      "center_classification_date": "20260722",
      "report_date": "20260729",
      "code_info": "UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)"
    }
  ]
}