      "status": "Terminated",
      "city": "Glamsbjerg",
      "state": "",
      "country": "Denmark",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74649",
      "recalling_firm": "Innovision A/S",
      "address_1": "Skovvaenget 2",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, to Massachusetts and Washington, DC; and, the countries of Canada, Germany, and United Kingdom.",
      "recall_number": "Z-2769-2016",
      "product_description": "Innocor Cardiopulmonary Exercise and Lung Function Testing System      Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.",
      "product_quantity": "8",
      "reason_for_recall": "The purpose of the recall is to replace data transmission cable (USB-cable) connecting the lnnocor device with an external computer.    The replacement USB cable will provide galvanic isolation of the lnnocor from the external computer via opto-couplers, as opposed to the grounding protection currently in use.    The replacement is intended to prevent risk of electric shock.",
      "recall_initiation_date": "20160513",
      "center_classification_date": "20160908",
      "termination_date": "20170210",
      "report_date": "20160914",
      "code_info": "Ref. INN00400, Serial no. 1551369, 1540362, 1540363, and 1536361.",
      "more_code_info": ""
    }
  ]
}