  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianola",
      "address_1": "1 Bayer Dr",
      "reason_for_recall": "CertegraWorkstation when used in conjunction with the Medrad Stellant CT Injection System may exhibit thermal events, such as: smoke and sparks of fire.",
      "address_2": "",
      "product_quantity": "165 systems",
      "code_info": "Serial Numbers:   KSA0517595  KSA0518535  KSA0519948  KSA0518539  KSA0518498  KSA0517618  KSA0518500D  KSA0518536  KSA0518456  KSA0518504  KSA0518520  KSA0519940  KSA0518507  KSA0519991  KSA0517568  KSA0517592  KSA0517649  KSA0518496  KSA0518508  KSA0519971  KSA0518534  KSA0517600  KSA0517566  KSA0517587  KSA0518469  KSA0518452  KSA0518460  KSA0518532  KSA0518516  KSA0519946  KSA0517641  KSA0519968  KSA0519941  KSA0518399  KSA0518479  KSA0517609  KSA0517617  KSA0518446  KSA0518453  KSA0518410  KSA0518466  KSA0517581  KSA0517574  KSA0518524  KSA0519956  KSA0518459  KSA0518495  KSA0518481  KSA0518513  KSA0518528  KSA0519989  KSA0518429  KSA0518523  KSA0517620  KSA0518509  KSA0517634  KSA0519964  KSA0518418  KSA0518474  KSA0518526  KSA0518468  KSA0518499  KSA0518463  KSA0517571  KSA0517648  KSA0517564  KSA0517644  KSA0518423  KSA0517582  KSA0518427  KSA0518518  KSA0518419  KSA0518397  KSA0518505  KSA0518522  KSA0517660  KSA0518492  KSA0519955  KSA0517629  KSA0518461  KSA0519951  KSA0518471  KSA0517597  KSA0517565  KSA0517656  KSA0518506  KSA0517607  KSA0517659  KSA0518491  KSA0517632  KSA0519978  KSA0517590  KSA0517612  KSA0517613  KSA0518447  KSA0518424  KSA0518400  KSA0519939  KSA0518431  KSA0518435  KSA0518515  KSA0518490  KSA0517624  KSA0518455  KSA0517661  KSA0518493  KSA0519945  KSA0517575  KSA0517588  KSA0517616  KSA0517653  KSA0517657  KSA0517658  KSA0518449  KSA0519982  KSA0519959  KSA0518426  KSA0519984  KSA0519976  KSA0519952  KSA0518511  KSA0518540  KSA0517584  KSA0517651  KSA0519947  KSA0518413  KSA0518405  KSA0518502  KSA0519954  KSA0519980  KSA0518531  KSA0518443  KSA0518529  KSA0519972  KSA0518510  KSA0517637  KSA0519990  KSA0518484  KSA0517602  KSA0518437  KSA0518477  KSA0517583  KSA0518530  KSA0518412  KSA0517577  KSA0519958  KSA0518442  KSA0518525  KSA0517635  KSA0517662  KSA0519957  KSA0519961  KSA0519981  KSA0518521  KSA0517646  KSA0518464  KSA0517576  KSA0518420  KSA0518433  KSA0519960  KSA0518414  KSA0517596  KSA0519970  KSA0518473",
      "center_classification_date": "20160908",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to United Arab Emirates, Australia, Canada, Germany, Spain, and United Kingdom.",
      "state": "PA",
      "product_description": "Certegra Workstation used in conjunction with Medrad Stellant CT Injection System   Catalog Number # CWKS DISPLAY  Serial Numbers   The Medrad Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications",
      "report_date": "20160914",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bayer Healthcare",
      "recall_number": "Z-2768-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74974",
      "termination_date": "20161209",
      "more_code_info": "",
      "recall_initiation_date": "20160802",
      "postal_code": "15051-9702",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}