  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Canton",
      "address_1": "5449 Research Dr",
      "reason_for_recall": "Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges.  An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.",
      "address_2": "",
      "product_quantity": "79 kits",
      "code_info": "Catalog number: HC-DIL  Lot:  072414-206  Expiration date:  2016-07",
      "center_classification_date": "20150915",
      "distribution_pattern": "US Nationwide Distribution.",
      "state": "MI",
      "product_description": "MedTest DX Control Reconstitution Fluid  Kit configuration:  20 x 5 ml.  Bottle:  5 ml amber glass  Cap:  18 mm white seal plug seal rubber  HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200.  in vitro diagnostic.",
      "report_date": "20150923",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "MedtestDx, Inc.",
      "recall_number": "Z-2768-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "71993",
      "termination_date": "20160523",
      "more_code_info": "",
      "recall_initiation_date": "20150515",
      "postal_code": "48188-2261",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}