  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "250 E 96th St",
      "reason_for_recall": "Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.",
      "address_2": "ste 125",
      "product_quantity": "6 units",
      "code_info": "Model number NN-7000   SN: 00231, 00232, 00242, 00245, 00248, 00250",
      "center_classification_date": "20150915",
      "distribution_pattern": "US Distribution to states of: MD, IN, CA, and NC.",
      "state": "IN",
      "product_description": "NICO Myriad Console  Model number NN-7000     The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece.     For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.",
      "report_date": "20150923",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Nico Corp.",
      "recall_number": "Z-2767-2015",
      "initial_firm_notification": "FAX",
      "product_type": "Devices",
      "event_id": "72015",
      "termination_date": "20150922",
      "more_code_info": "",
      "recall_initiation_date": "20150729",
      "postal_code": "46240-3781",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}