{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83574",
      "recalling_firm": "Roche Diabetes Care, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution.",
      "recall_number": "Z-2766-2019",
      "product_description": "Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.",
      "product_quantity": "N/A",
      "reason_for_recall": "Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).",
      "recall_initiation_date": "20170612",
      "center_classification_date": "20190930",
      "termination_date": "20201030",
      "report_date": "20191009",
      "code_info": "Serial numbers less than xxx25525056    Updated serial numbers on 10/12/2019:   Performa II: Between XXX20000000 and XXX25525056    Updated 10/18/2019:    Performa II:  from XXX20000000 to XXX25525056"
    }
  ]
}