{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Maple Grove",
      "address_1": "1 Scimed Pl",
      "reason_for_recall": "Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preventing subsequent sealing of the valve when desired. This Field Correction reinforces the Directions for Use (DFU) and gives further guidance on correct use of the hemostasis valve to avoid cross-threading and to securely seal the valve, minimizing potential for undesired blood leakage.",
      "address_2": "",
      "product_quantity": "29,169 units total",
      "code_info": "15611295  15624432  15653185  15657827  15675111  15640386  15679748  15684667  15698742  15716330  15736494  15736689  15746782  15711284  15753133  15757057  15760114  15764019  15767450  15768507  15770114  15772280  16635566  16691146  17148680  17628922  16675497  16723382  16731675  16738582  16740109  16752661  16760782  16880335  16897923  16904754  16751834  16893570  16894795  16933430  16937185  16940509  16943256  17038874  17041892  17044130  17046074  17123055  17126755  17127496  17133735  17138028  17206369  17210088  17211902  17214752  17218900  17296059  17300015  17388374  17388375  17388376  17388377  17453643  17453644  17708567  17708568  17708569  17708570  18014966  18014967  18014968  18075932  18075933  18075934  18075935  18110488  18204409  18204410  18204411  F00548801  F00548802  F00548803  F00548804  F00548805  F00548806  F00548807  F00548808  F00548809  F00548810  F00548811  F00548812  F00548813  F00548814  F00548815  F00548816  F00548817  F00548818  F00548819  F00548820  F00548821  F00548901  F00548902  F00548903  F00548904  F00548905  F00548906  F00548907  F00548908  F00548909  F00548910  F00548911  F00548912  F00548913  F00548914  F00548915  F00548916  F00548917  F00548918  F00548919  F00548920  F00548921  F00549001  F00549002  F00549003  F00549004  F00549005  F00549006  F00549007  F00549008  F00549009  F00549010  F00549011  F00549012  F00549013  F00549014  F00549015  F00549016  F00549017  F00549018  F00549019  F00549101  F00549102  F00549103  F00549104  F00549105  F00549106  F00549107  F00549108  F00549109  F00549110  F00549111  F00549112  F00549113  F00549114  F00549115  F00549116  F00549117  F00549118  F00549119  F00549120  F00549121  F00549122  F00549123  F00549124  F00549125  F00549126  F00549127  F00549128  F00549130  F00550301  F00550302  F00550303  F00550304  F00550305  F00550306  F00550307  F00550308  F00550309  F00550310  F00550311  F00550312  F00550313  F00550314  F00550315  F00550316  F00550317  F00550318  F00550319  F00550320  F00550321  F00550401  F00550402  F00550403  F00550404  F00550405  F00550406  F00550407  F00550408  F00550409  F00550410  F00550411  F00550412  F00550413  F00550414  F00550415  F00550416  F00550417  F00550418  F00550419  F00550420  F00550421  F00550422  F00550423  F00550501  F00550502  F00550503  F00550504  F00550505  F00550506  F00550507  F00550508  F00550509  F00550510  F00550511  F00550512  F00550513  F00550514  F00550515  F00550516  F00550517  F00550518  F00550601  F00550602  F00550603  F00550604  F00550605  F00550606  F00550607  F00550608  F00550609  F00550610  F00550611  F00550612  F00550613  F00550614  F00550615  F00550616  F00550617  F00550618  F00550619  F00550620  F00550621  F00550622  F00550623  F00551701  F00551702  F00551703  F00551704  F00551705  F00551706  F00551707  F00551708  F00551709  F00551710  F00551711  F00551712  F00551713  F00551714  F00551715  F00551716  F00551717  F00551718  F00551719  F00551720  F00551721  F00551722  F00551801  F00551802  F00551803  F00551804  F00551805  F00551806  F00551807  F00551808  F00551809  F00551810  F00551811  F00551812  F00551813  F00551814  F00551815  F00551816  F00551817  F00551818  F00551819  F00551820  F00551901  F00551902  F00551903  F00551904  F00551905  F00551906  F00551907  F00551908  F00551909  F00551910  F00551911  F00551912  F00551913  F00551914  F00551915  F00551916  F00551917  F00551918  F00551919  F00552001  F00552002  F00552003  F00552004  F00552005  F00552006  F00552007  F00552008  F00552009  F00552010  F00552011  F00552012  F00552013  F00552014  F00552015  F00552016  F00552017  F00552018  F00552019  F00552020  F00552021  F00553501  F00553502  F00553503  F00553504  F00553505  F00553506  F00553507  F00553508  F00553509  F00553510  F00553511  F00553512  F00553513  F00553514  F00553515  F00553516  F00553517  F00553518  F00553519  F00553520  F00553521  F00553522  F00553523  F00553524  F00553525  F00553526  F00553527  F00553528  F00553601  F00553602  F00553603  F00553604  F00553605  F00553606  F00553607  F00553608  F00553609  F00553610  F00553611  F00553612  F00553613  F00553614  F00553615  F00553616  F00553617  F00553618  F00553619  F00553620  F00553621  F00553622  F00553623  F00553701  F00553702  F00553703  F00553704  F00553705  F00553706  F00553707  F00553708  F00553709  F00553710  F00553711  F00553712  F00553713  F00553714  F00553715  F00553716  F00553717  15523924  15534295  15547790  15547791  15547792  15549844  15557261  15557262  15569373  15573927  15558971  15577456  15577564  15577565  15582831  15591639  15596436  15597413  15604737  15607853  15610775  15614197  15614198  15630136  15624240  15652977  15640384  15657830  15660313  15640385  15647162  15630139  15675108  15675110  15676735  15647163  15679745  15679746  15691631  15684861  15691634  15775097  15778019  15780647  15788130  15791345  15795851  15799915  15802493  15818893  15828834  15837151  17524411  17524416  17524417  17524418  17524720  17524721  15846553  15851186  15854603  15858077  15862218  15866281  15870175  15876700  15883165  15889807  15892273  15895940  15899488  15904004  15910338  15913991  17524722  17524723  17524724  17524725  17524726  15917331  15921281  15923967  15931042  15934214  15938318  15945693  15955251  15965197  15968027  15972749  15976621  15980066  15982571  17524727  17524729  15986987  15989302  15992406  15992934  15997550  16005081  16008695  16015248  16018005  16022224  16025224  16031641  16039474  16042993  16045736  16049761  16052878  17524731  16056581  16059512  16063042  16072486  16075309  16075311  16078702  16085500  16089346  16092734  16095740  16099082  16102078  16102676  16107064  16110480  16113541  16116554  16119788  16125460  16128359  16131915  16138326  16141222  16144572  16145326  16150818  16155310  16158711  16161518  16168782  16176187  16183060  16186299  16193942  16196930  16200326  16204070  16207559  16210178  16213052  16218852  16222332  16225384  16228241  16231880  16235347  16239221  16245797  16249323  16253202  16256863  16262989  16266635  16270213  16273633  16282152  16284577  16287159  16289074  16294255  16297274  16300507  16311011  16311108  16316943  16319992  16323506  16326972  16330057  16334237  16338001  16341153  17524732  16344102  16347281  16410911  16418478  16422805  16423443  16455414  16483226  16519540  16519541  16547565  16548609  17524733  16579911  16583156  16590621  16598142  16601174  16602910  16608075  16611632  16615627  16619147  16621806  16623957  16626169  16629592  16638766  16642109  16642111  16649392  16654987  16655276  16655277  16655278  16678438  16679009  16682438  16695334  16698285  16701585  16702550  17524734  17524735  16709414  16713086  16716476  16719767  16719776  16727778  16735188  16738581  16748485  16752079  16768234  16772083  16775461  17524736  17538718  16787947  16790633  16799406  16803525  16805226  16807376  16812947  16816972  16820960  16823548  16826394  16830360  16833974  16849203  16850145  16852559  16857214  16860704  16863193  16865854  16867902  16870755  16875093  16875734  16900812  16911587  16915195  16918019  16922572  16927174  16946167  16949828  16953116  16957400  16957818  16964405  16967574  16970995  16974815  16979304  16984104  16990814  16994018  16999911  17003660  17006537  17010728  17013673  17020297  17023538  17027982  17032220  17035414  17046073  17053900  17056855  17058566  17064562  17068255  17071798  17075340  17078716  17083015  17086912  17088543  17091394  17092084  17101486  17103308  17107951  17111599  17115175  17118418  17142698  17148056  17152068  17155707  17158556  17161832  17164904  17167875  17174211  17175418  17179248  17182780  17185582  17190030  17193308  17199428  17225603  17227800  17232654  17234871  17235440  17240940  17249477  17253521  17256408  17263096  17268930  17272646  17276355  17280717  17284570  17288172  17291762  17303754  17313347  17313348  17313349  17313350  17332902  17332903  17332904  17332905  17332906  17332907  17332908  17353307  17353308  17353309  17408325  17408326  17408327  17408328  17408329  17425072  17425073  17425074  17425075  17425076  17475599  17475840  17475841  17475842  17501056  17501057  17501058  17501059  17501080  17531156  17531157  17531158  17531159  17548469  17548470  17548471  17548472  17548473  17590049  17590050  17590051  17590052  17590053  17600427  17600428  17600429  17600430  17600431  17647220  17647221  17647222  17647223  17647224  17665914  17665915  17665916  17665917  17665918  17692714  17692717  17692718  17692719  17692880  17739251  17739252  17739253  17739254  17739255  17755161  17755162  17755163  17755164  17755165  17778453  17778454  17778455  17778456  17778457  17812333  17812334  17812335  17812336  17812337  17852727  17852728  17852729  17852730  17852731  17867817  17867818  17867819  17867920  17867921  17901058  17901059  17901080  17901081  17901082  17925122  17925123  17925124  17925125  17925126  17952966  17952967  17952968  17952969  17952970  17960718  17960719  17978215  17978216  17978217  18036339  18036460  18036461  18036462  18036463  18036467  18036468  18036469  18036470  18036471  18036472  18075936  18075937  18075938  18110489  18115034  18115035  18120793  18154221  18154222  18158642  18158643  18158644  18158645  18158646  18204079  18204400  18204401  18233257  18233259  18233260  18233261  18233262  18233263  16355622  16358726  16363048  16367842  16372912  16379808  16384696  16390596  16393857  16397335  16401304  16422058  16426378  16429530  16432658  16436681  16440016  16443231  16446666  16457879  16465045  16475632  16482045  16489282  16491854  16500488  16510368  16521529  16547562  16590794  16593999  17797959  17818364  17900405  18029494  18085990  18085991  18085992  18085993  18085994  18091074  18091075  18091076  18097802  18097803  18097804  18111728  18179106  18179107  18179108  18179109  18179110  18179111  18204405  18204406  18204407  18252945  18252947  18252948  16751831  16893569  16943257  17046075  17138029  17218901  17300014  17313351  17453642  15711285  15722505  15716329  15733228  15742846  15746783  15750090  15760116  15785032  15802490  15808197  15821967  15825660  15833585  15839839  17524103  17524104  17524105  17524106  17524107  17524108  15870735  15906798  17524109  17524404  15942740  15948999  15952034  15957983  15961293  17524113  17524115  17524118  17524400  17524402  17539282  17539283  15992933  16028407  16035580  17538716  17538719  16135267  16145323  16164714  16168810  16172942  16179406  16242184  16256223  16277475  16409463  16422803  16423442  16576447  16627421  16691145  16702549  16706056  17538717  17539281  16719766  16743849  16764290  16773649  16773650  16794371  16794372  16840109  16842573  16852558  16865853  16875735  16930186  16983795  16987369  17014384  17170094  17246303  17265821  17313481  18158641  16257502  16404896  16450763  16454321  16461294  16471886  16479343  16485513  16495598  16498932  16506317  16513845  16517321  16524449  16528428  16531473  16535543  16539323  16543352  17797953  17818363  18029490  18111727  18204403  18252949",
      "center_classification_date": "20150915",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Oman, Panama, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, White Russia.",
      "state": "MN",
      "product_description": "Boston Scientific,  WATCHMAN Access System  Access Sheath with Dilator, 14F Single curve( M635TU10060, TU1006) , 14F Double curve (M635TU20060, TU2006) , & 14F Anterior curve (M635TU40060, TU4006).  For cardiac use.",
      "report_date": "20150923",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Boston Scientific Corporation",
      "recall_number": "Z-2766-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72045",
      "termination_date": "20170404",
      "more_code_info": "",
      "recall_initiation_date": "20150825",
      "postal_code": "55311-1565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}