  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Marlborough",
      "address_1": "250 Campus Dr",
      "reason_for_recall": "Tip Breakage at the glue joint between the obturator body and tip",
      "address_2": "",
      "product_quantity": "1440 units",
      "code_info": "Lot Codes:  512501, 512502, 512504, 512588, 601501, 601535, 601536, 601587, 602501, 602502, 602503, 602504, 603522, 604557, 604558, 604591, 605501, 605502, 605543, 605547.",
      "center_classification_date": "20160908",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "MA",
      "product_description": "SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI.  ILS-0914-20-OB (9G Introducer Localization Set (OB Model) for ATEC0914-20MR Handpiece)",
      "report_date": "20160914",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hologic, Inc",
      "recall_number": "Z-2765-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74839",
      "termination_date": "20170330",
      "more_code_info": "",
      "recall_initiation_date": "20160802",
      "postal_code": "01752-3020",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}