  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Maple Grove",
      "address_1": "1 Scimed Pl",
      "reason_for_recall": "Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preventing subsequent sealing of the valve when desired. This Field Correction reinforces the Directions for Use (DFU) and gives further guidance on correct use of the hemostasis valve to avoid cross-threading and to securely seal the valve, minimizing potential for undesired blood leakage.",
      "address_2": "",
      "product_quantity": "29,169 units total",
      "code_info": "16752208  16891755  17138030  17211903  17214753  17296669  17395755  17483898  17712231  18053939  18120336  F00541702  F00541703  F00541707  F00541708  F00541709  F00541903  F00541904  F00541905  F00541906  F00541907  F00541908  F00541909  F00541913  F00541914  F00541916  F00544503  F00544504  F00544505  F00544508  F00544509  F00544511  F00544703  F00544704  F00544706  F00544707  F00544709  F00544710  F00545201  F00547501  F00547502  F00547505  F00547506  F00547507  F00547508  F00547509  F00547510  F00547511  F00547512  F00547514  F00548202  F00548204  F00548205  F00548206  F00548207  F00548210  F00548211  F00548213  F00548215  15513009  15556842  15559607  15559608  15573287  15576411  15579547  15583022  15586521  15597796  15608021  15634208  15639752  15645828  15648179  15650910  15664113  15688025  15694584  15711219  15719185  15722025  15746165  15760115  15767309  15791346  15798829  15802896  15802971  17514543  15846552  15851185  15859123  15870174  15883167  15889808  15906797  15913993  15918189  17505941  15921282  15927421  15931041  15934213  15935418  15942659  15955253  15957982  15965734  15972756  15980369  17514542  17516996  17517162  17517166  15989303  15990230  15992405  15996022  16018190  16022222  16039472  16063043  16069159  16081852  16095359  16102740  16125461  16135272  16138323  16151320  16168784  16172940  16180118  16190833  16208087  16228219  16239134  16242466  16245796  16271044  16274328  16284579  16297407  16326970  16330055  16337479  16342271  17505943  17522735  16410910  16457878  16459272  16511473  16528429  16535544  16547561  16566477  16586724  16601179  16605567  16611630  16615630  16626168  16669688  16676235  16684501  16691147  16706057  17505944  16709410  16714136  16719785  16743850  16772082  16783761  16794223  16799658  16807375  16811004  16812948  16820799  16826393  16835331  16858003  16860706  16875092  16919733  16927173  16931941  16946166  16957819  16984103  17015190  17016764  17031819  17053901  17088542  17588070  17158555  17161831  17162162  17162163  17162164  17174212  17190029  17193307  17256409  17265820  17271026  17274694  17281586  17282611  17308887  17331530  17337336  17368985  17380187  17384737  17384738  17384739  17385000  17389788  17400531  17404701  17408444  17421244  17424065  17428086  17442611  17462354  17544822  17553573  17563836  17566509  17566874  17566875  17566876  17576385  17584629  17607405  17656415  17665904  17666623  17674322  17674431  17680100  17750412  17801348  17830899  17845077  17855684  17881675  17885212  17905780  17955965  17969721  17989462  17994393  18059017  18109305  18139934  18196864  18212340  18224199  18273908  18279543  16358559  16372911  16401305  16422059  16429531  16440015  16465046  16482043  16489283  16500682  16507541  16510365  16521530  16579912  16594442  17797950  18029497  18114255  18135004  18149586  18264433  18273909  16752207  16893571  16943258  17049930  17133736  17214754  17289126  17296670  17394782  17483899  17715339  18015413  18053936  18104653  F00543403  F00543404  F00543409  F00543410  F00543411  F00544801  F00544802  F00544804  F00544805  F00544806  F00544901  F00544902  F00544903  F00544904  F00544905  F00546002  F00546004  F00546005  F00546006  F00546007  F00546008  F00546011  F00546012  F00547702  F00547704  F00547705  F00547706  F00547710  F00547711  F00547712  F00547713  F00547715  F00548001  F00548003  F00548004  F00548005  F00548006  F00548007  F00548008  F00548009  F00548010  F00548011  F00548012  F00548014  F00548015  15533157  15533159  15534567  15560537  15560538  15563850  15584346  15601788  15613894  15624071  15627321  15630804  15639756  15645827  15650908  15656641  15660122  15664112  15670042  15675296  15678800  15694583  15712768  17510339  17513899  17514330  15725135  15732982  15752943  15757059  15761457  15764438  15767310  15768509  17510421  17514339  17514544  15785030  15802897  15828835  15837276  15839943  15873851  15880319  15892274  15904005  15905274  15907468  17514540  15917731  15923968  15927422  15942658  15945694  15952033  15972750  15980371  15982572  15986988  15996021  16008893  16009246  16012590  16025225  16028569  16031642  16049990  16050155  16053803  16065499  16075310  16095358  16110279  16113543  16128522  16135270  16138985  16141220  16155838  16161519  16179407  16187284  16196931  16201410  16204071  17516997  16210174  16225593  16231399  16239135  16245795  16249321  16260101  16264696  16274329  17516991  16282440  16284578  16294524  16304635  16316948  16326971  16338000  16342273  17517172  17517421  17517701  16405297  16408162  16436680  16443230  16447292  16457877  16511475  16524446  16529379  16543353  16551893  16551899  16559456  16566248  16619148  16621808  16623958  16635565  16638767  16649394  16654988  16675496  16676236  16684502  16691148  17588071  16701580  17504979  17505620  17505627  17522737  16716474  16723383  16735187  16740111  16752662  16764291  16775462  16790634  16797991  16799659  16803524  16805225  16812946  16823549  16842574  16857215  16858004  16897924  16900811  16902464  16943540  16953117  16967573  16987245  16999910  17003018  17015191  17020296  17035415  17056856  17075341  17091393  17101485  17139718  17142697  17164903  17167874  17175419  17199427  17232655  17246388  17268931  17308289  17323937  17347993  17360281  17400535  17404703  17426248  17445090  17462355  17513733  17547678  17549055  17562117  17563262  17567816  17741851  17585954  17598806  17607108  17628229  17646696  17662280  17673314  17684461  17687772  17695516  17734422  17734431  17757446  17759451  17780348  17801349  17815706  17847860  17852577  17861964  17870082  17899546  17923728  17935420  17952163  17960377  17981460  18000250  18070820  18187096  18192386  18212343  18232577  16355621  16379807  16426377  16471884  16478927  16489284  16491855  16498933  16507540  16517322  16586726  16590623  16594441  16608077  16615629  16619150  17820313  18029498  18095185  18109304  18133924  18135005  18149585  18232578  18243072  16753120  16897922  16933432  16937186  17127497  17206370  17210497  17217173  17218902  17227099  17395756  17484720  17720340  17725256  18049416  18120335  18124913  F00544001  F00544102  F00544103  F00544104  F00544105  F00544109  F00544110  F00544111  F00544602  F00544603  F00544604  F00544605  F00544609  F00544610  F00544611  F00545001  F00545006  F00545007  F00545009  F00545010  F00545011  F00545012  F00545401  F00545403  F00545406  F00545410  F00545411  F00545412  F00547001  F00547002  F00547004  F00547005  F00547006  F00547009  F00547101  F00547102  F00547103  F00547104  F00547105  F00547106  F00547107  F00547108  F00547109  F00547110  F00547112  F00547113  F00547114  F00547115  F00547603  F00547604  F00547606  F00547607  F00547609  F00547610  F00547611  F00547612  F00547613  F00547614  F00547615  F00547801  F00547802  F00547803  F00547804  F00547805  F00547808  F00547809  F00547810  F00547811  F00547812  F00547813  F00547814  F00547815  F00547901  F00547902  F00547903  F00547904  F00547905  F00547906  F00547907  F00547908  F00547909  F00547910  F00547911  F00547912  F00547913  F00547914  F00547915  F00548302  F00548303  F00548304  F00548305  F00548306  F00548309  F00548310  F00548311  F00548312  F00548313  F00548314  15534565  15534566  15560536  15567410  15573288  15601787  15610674  15613896  15617630  15617819  15641459  15624074  15649001  15656642  15659939  15663165  15670041  15675294  15691175  15704329  17506121  17506122  17510420  15722027  15726314  15726790  15735937  15750089  15757080  15758621  17593205  15768508  17489860  17489861  17489862  15780922  15788129  15795135  15802898  15825661  15833586  15843113  15854664  15862600  15876340  15907469  15911364  15913992  15931040  15945696  15955252  15961295  15968028  15976620  15980375  17510423  17513885  17517870  17517873  15992404  16000476  16008894  16014993  16025226  16031644  16039471  16043831  16049993  16050156  16053802  16072487  16078703  16099084  16102961  16116555  16125462  16131916  16138324  16144976  16148062  16155839  16158712  16165648  16168783  16176189  16190832  16194169  16196932  17505633  17517874  17522736  16210175  16213053  16218853  16239132  16245794  16253201  16256395  16271046  16277162  17514334  16289073  16291559  16311107  16324449  16330056  16341152  16342274  17516999  17522729  16414731  16450765  16511476  16524447  16525613  16529939  16531474  16539324  16552003  16559519  16567395  17517704  16576443  16583155  16591524  16591525  16599033  16599034  16599035  16604972  16604974  16621809  16642110  16642112  16649393  16676237  16684503  16698601  16701581  17517706  17517707  16709412  16716475  16731674  16738583  16748486  16759988  16768235  16778857  17505628  17523740  16794222  16799940  16803521  16805224  16807374  16820798  16830259  16838592  16845891  16858005  16867900  16870754  16890153  16918018  16922573  16930291  16957199  16970994  16974814  16990338  17006538  17010685  17015193  17023537  17058567  17060697  17072362  17078413  17083017  17086914  17111598  17124358  17148055  17170093  17179249  17246304  17261865  17280718  17288434  17303753  17320842  17335800  17345077  17355465  17361618  17367565  17385002  17400534  17404702  17431206  17440837  17456606  17505566  17513728  17549051  17549052  17557930  17563433  17563837  17742337  17572195  17594047  17611275  17623145  17638865  17684460  17687771  17700105  17743276  17750413  17776215  17789857  17795361  17810105  17820605  17829287  17845078  17852460  17861963  17866075  17894268  17910306  17935421  17951926  17960378  18079952  18139617  18142003  18196863  18212341  18220251  18224620  18255944  18284349  16370230  16388172  16401306  16426379  16432657  16440017  16447294  16461296  16471885  16478928  16485510  16491856  16495599  16502705  16510366  16513846  16566478  16576449  17820317  18029601  18085143  18104652  18129620  18135006  18153570  18170973  18201250  18239268  18243071  18264434  18284348  18288690  18293133  18297522  18304018  16751839  16893572  16937642  17038873  17123056  17126756  17210498  17395757  17484721  17729577  18054580  18109303  18124912  18268763  F00540901  F00540903  F00540909  F00540910  F00540911  F00545502  F00545506  F00545507  F00545508  F00545511  F00545601  F00545602  F00545604  F00545605  F00545606  F00545607  F00545609  F00545611  F00545612  F00545712  F00545808  F00545809  F00546108  F00546301  F00546303  F00546304  F00546306  F00546307  F00546501  F00546502  F00546505  F00546506  F00546507  F00546509  F00546510  F00546701  F00546702  F00546704  F00546706  F00548501  F00548502  F00548506  F00548507  F00548508  F00548509  F00548510  F00548601  F00548602  F00548603  F00548604  F00548605  F00548606  F00548607  F00548608  F00548609  F00548610  F00548701  F00548702  F00548704  F00548705  F00548706  F00548708  F00548709  F00548710  F00549601  F00549602  F00549603  F00549605  F00549606  F00549607  F00549608  F00549609  15533414  15534507  15554778  15559112  15567409  15576410  15513004  15583465  15586522  15596777  15597797  15597799  15601789  15634207  15636698  15641980  17510336  15639840  15691176  15704330  15709031  17510338  15726789  15742845  15768979  15775654  15780644  15785551  15866546  15899484  15904700  15921283  15938317  15945695  15961294  15976299  15980377  15980378  15982757  16005082  16018189  16035581  16043830  16049994  16066321  16069158  16092736  16099085  16119789  16128521  16131917  16144975  16150819  16172941  17587903  16176188  16194170  16200327  17489863  17489864  16213051  16218855  16235346  16253200  16256396  16260102  16274330  16277970  17523742  16282535  16301056  16304511  16311248  16317993  16319993  16334234  16342277  17512673  17517702  16418477  16552008  16605571  16611631  16621807  16629591  16645865  16649395  16676238  16698284  17485956  16713085  16727779  16740110  16773087  16797992  16799941  16816970  16833973  16835333  16863192  16879810  16883020  16911586  16915196  16930188  16931943  16964404  16979303  17013672  17015194  17049932  17064564  17068256  17103307  17125140  17138031  17152067  17155705  17182781  17235441  17240941  17249478  17274686  17274689  17284571  17315901  17327649  17418948  17435610  17449151  17474986  17505149  17505151  17512095  17516438  17555483  17563263  17563434  17580530  17598807  17607463  17628583  17643288  17766401  17785911  17829286  17838466  17858126  17866074  17881676  17904999  17914603  17935423  17969722  17989461  17994394  18005577  18059018  18075560  18164716  18187095  18224198  18268761  16359060  16379809  16390594  16429529  16438600  16454320  16461295  16475631  16485511  16489285  16495700  16506318  16513844  16521531  16528430  16535546  16547564  16562258  16569781  16583154  16590622  16598143  16601175  17797951  18029602  18100401  18124914  18144934  18157433  18164718  18268762  16752418  16893573  16933431  17041893  17042602  17112228  17118419  17478392  17731691  18054581  18129539  18279544  F00540702  F00540708  F00540710  F00543803  F00543807  F00543808  F00543810  F00543811  F00544202  F00544203  F00544205  F00544206  F00544207  F00544208  F00544210  F00544211  F00544212  F00544213  F00544214  F00545901  F00545903  F00545904  F00545906  F00545907  F00545909  F00545910  F00546201  F00546204  F00546207  F00546208  F00546401  F00546403  F00546404  F00546405  F00546407  F00546408  F00546409  F00546410  F00546605  F00546606  F00546607  F00546609  F00546803  F00546805  F00546901  F00546903  F00546904  F00546905  F00546906  F00546907  F00546909  F00546910  F00546911  F00546912  F00547201  F00547203  F00547204  F00547205  F00547206  F00547209  F00547210  F00547301  F00547302  F00547303  F00547304  F00547305  F00547306  F00547307  F00547308  F00547309  F00547310  F00547401  F00547404  F00547405  F00549501  F00549503  F00549504  F00549505  F00549506  F00549507  F00549508  F00549509  15533431  15555660  15556765  15558791  15576412  15579548  15586576  15608022  15610675  15637569  15650909  15678801  15684755  15697854  15711401  17510337  15732981  15739114  15770115  15772889  15775655  17513887  15917732  16022223  16045735  16072787  16081851  16102079  16102965  16119791  16141221  16148063  16165647  16179408  16187286  16204068  17517168  16208090  16210176  16218850  16242465  16249320  16271045  16282651  16291980  16307822  16319990  16334236  16342278  17512672  17517703  16421082  16524448  16605572  16626167  16638768  16638769  16642108  17588074  16655802  16655806  16676239  16678436  16684504  16691076  16706055  17505940  17517705  17523741  16709409  16719786  16743851  16755490  16787948  16797993  16799942  16811005  16816971  16835335  16845892  16861290  16886782  16898028  16904755  16909513  16909514  16949829  16983797  16983799  16994017  17027981  17060695  17060696  17078717  17091552  17107949  17152260  17176021  17185583  17214494  17214495  17214496  17214497  17214498  17214499  17272647  17276356  17282618  17286603  17286605  17286606  17286607  17289390  17289392  17305049  17320957  17320958  17320959  17332272  17344774  17345641  17368981  17368982  17368984  17385001  17396196  17400277  17421249  17442610  17462953  17462954  17468852  17468853  17474814  17479331  17480580  17480581  17480582  17486462  17486463  17492416  17492417  17510073  17510074  17513731  17513732  17532170  17532176  17544781  17547598  17553574  17563269  17582162  17582357  17582358  17585327  17600945  17603235  17607109  17608612  17650684  17804597  17678068  17684518  17695947  17695957  17708084  17713328  17767544  17767545  17767546  17767549  17767550  17784915  17820604  17830883  17830884  17830885  17833546  17858127  17859249  17880942  17914602  17924324  17924325  17924326  17948921  17955966  18010012  18014257  18157248  18177409  18192387  18205715  18216300  18280016  16353507  16404895  16446665  16450764  16465047  16482044  16485512  16495597  16502704  16510367  16513847  16539322  16547563  16562259  16586725  16594440  16601178  16608076  16611633  16619149  16623959  18029603  18085144  18100402  18153571  18280015",
      "center_classification_date": "20150915",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Oman, Panama, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, White Russia.",
      "state": "MN",
      "product_description": "Boston Scientific, WATCHMAN Left Atrial Appendage Closure Device with Delivery System, 21mm x 12F ( M635WU21060,  WU21060, 24mm x 12F(M635WU24060, WU2406) , 27mm x 12F(M635WU27060, WU2706) , 30mm x 12F(M635WU30060, WU3006)  , 33mm x 12F( M635WU33060, WU3306).  For cardiac use.",
      "report_date": "20150923",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Boston Scientific Corporation",
      "recall_number": "Z-2765-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72045",
      "termination_date": "20170404",
      "more_code_info": "",
      "recall_initiation_date": "20150825",
      "postal_code": "55311-1565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}