  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83163",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution -  US Nationwide and the countries of  Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.",
      "recall_number": "Z-2763-2019",
      "product_description": "The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755.  A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.",
      "product_quantity": "Access Cortisol Reagent Assay",
      "reason_for_recall": "Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).",
      "recall_initiation_date": "20180918",
      "center_classification_date": "20190930",
      "termination_date": "20240221",
      "report_date": "20191009",
      "code_info": "Lot Number 831755."
    }
  ]
}