{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85999",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2757-2020",
      "product_description": "SIEMENS Healthineers Uroskop Omnia max   Model: 10762473 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.",
      "product_quantity": "194 units",
      "reason_for_recall": "Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis",
      "recall_initiation_date": "20200108",
      "center_classification_date": "20200805",
      "termination_date": "20220621",
      "report_date": "20200812",
      "code_info": "Serial Numbers:  4557  4574  4571  4099  4074  4014  4627  4626  4578  4645  4316  4615  4056  4587  4049  4068  4305  4015  4543  4616  4553  4333  4063  4655  4556  4022  4521  4121  4325  4111  4510  4565  4322  4554  4301  4597  4509  4661  4036  4519  4520  4043  4024  4560  4629  4530  4646  4318  4618  4575  4304  4142  4300  4095  4146  4649  4059  4562  4501  4309  4548  4313  4119  4135  4069  4020  4032  4612  4512  4053  4663  4061  4590  4593  4642  4312  4528  4060  4051  4506  4025  4037  4635  4054  4042  4033  4544  4541  4566  4132  4067  4120  4613  4112  4335  4336  4072  4652  4055  4108  4559  4511  4540  4126  4129  4104  4307  4334  4073  4631  4582  4570  4561  4641  4052  4572  4047  4134  4605  4326  4529  4603  4568  4654  4532  4102  4130  4317  4505  4062  4594  4096  4585  4637  4085  4143  4516  4662  4086  4598  4118  4573  4101  4589  4100  4321  4558  4591  4105  4106  4513  4538  4098  4537  4306  4539  4066  4310  4651  4624  4314  4545  4552  4082  4664  4125  4328  4124  4636  4329  4600  4601  4602  4094  4114  4117  4136  4303  4319  4140  4141  4500  4531  4630  4622  4535  4580  4660  4330  4332  4133  4588  4609",
      "more_code_info": ""
    }
  ]
}