  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irving",
      "address_1": "1921 Hurd Dr",
      "reason_for_recall": "The firm identified customer sites where modules which the safety pacemaker label is missing. Information regarding the label included in the Operations Manual has been updated to specify: 1) the risk of malfunction applies to anyone with a pacemaker who may work on these modules; 2) the minimum safety distance has increased to 200 mm (7.87\") from transport mechanisms. Any person with a pacemaker must not get closer than this distance from modules axes.",
      "address_2": "",
      "product_quantity": "159 devices",
      "code_info": "All Lot Numbers    ABT018 ABT070 ABT107 ABT135  ABT019 ABT074 ABT108 ABT136  ABT023 ABT075 ABT109 ABT137  ABT024 ABT077 ABT110 ABT138  ABT026 ABT078 ABT111 ABT139  ABT027 ABT081 ABT112 ABT140  ABT028 ABT082 ABT113 ABT142  ABT034 ABT083 ABT114 ABT143  ABT036 ABT084 ABT116 ABT144  ABT037 ABT086 ABT117 ABT145  ABT041 ABT088 ABT118 ABT146  ABT042 ABT089 ABT119 ABT147  ABT047 ABT091 ABT120 ABT149  ABT049 ABT092 ABT121 ABT150  ABT050 ABT093 ABT122 ABT151  ABT051 ABT094 ABT123 ABT152  ABT054 ABT096 ABT124 ABT153  ABT058 ABT097 ABT125 ABT154  ABT059 ABT098 ABT126 ABT156  ABT062 ABT099 ABT127 ABT157  ABT063 ABT100 ABT128 ABT158  ABT065 ABT101 ABT129 ABT159  ABT066 ABT102 ABT130  ABT067 ABT104 ABT131  ABT068 ABT105 ABT132  ABT069 ABT106 ABT133",
      "center_classification_date": "20190926",
      "distribution_pattern": "US: TX, VA, AZ, NV, MA, LA, MN, PA, SC, NJ, NY, CA, IL, IN, WA, FL, TN, PA, KS, ND, NC, CT    OUS: Argentina, Australia, Belgium, Canada, China, Denmark, France, Greece, Hong Kong, Ireland, Italy, New Zealand, Norway, Oman, Pakistan, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, UAE, UK,  Venezuela",
      "state": "TX",
      "product_description": "APS Accelerator modular systems, 208-230 V-, 50/60 Hz, 1800 VA, IVD, RX, Class 2 Laser Product,",
      "report_date": "20191002",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories, Inc",
      "recall_number": "Z-2755-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83651",
      "termination_date": "20210202",
      "more_code_info": "",
      "recall_initiation_date": "20180907",
      "postal_code": "75038-4313",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}