  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Grove City",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99210",
      "recalling_firm": "Tosoh Bioscience Inc",
      "address_1": "3600 Gantz Rd",
      "address_2": "",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Global Distribution. US Nationwide.",
      "recall_number": "Z-2746-2026",
      "product_description": "ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204",
      "product_quantity": "310 units",
      "reason_for_recall": "The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.",
      "recall_initiation_date": "20260612",
      "center_classification_date": "20260716",
      "report_date": "20260722",
      "code_info": "UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408",