{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arras CEDEX",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85693",
      "recalling_firm": "Normand-Info S.A.S.U.",
      "address_1": "7 rue Frederic Degeorge",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including states of: CA, GA, IL, IN, KS, MI, MO, NY, NC, OH, PA, SD, TN, VA, and WV;  and countries of: Bahrain, Belgium, Canada, China, Czech Republic, France, Germany, Hong Kong, Israel, Malaysia, Netherlands, Switzerland, Taiwan,Turkey, and United Kingdom.",
      "recall_number": "Z-2743-2020",
      "product_description": "Remisol Advance running Software Version 1.12    The Normand Info Remisol Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS &). It can also be connected to a host system.",
      "product_quantity": "663",
      "reason_for_recall": "Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.",
      "recall_initiation_date": "20200430",
      "center_classification_date": "20200731",
      "termination_date": "20220210",
      "report_date": "20200812",
      "code_info": "All serial numbers",
      "more_code_info": ""
    }
  ]
}