  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Siemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL respectively.",
      "address_2": "PO BOX 6101",
      "product_quantity": "Total number Dimension Vista - 32,165,  Dimension - 238,634",
      "code_info": "Dimension CREA Lot Numbers:   DB5195  FA5211  EA5233  DA5258  FA5275  BA5293  EA5315  EA5316  GA5342  FA5351  FA5352   FA6006  BA6026  BA6027  EB6047  GB6076  GA6097  BA6111  BA6112  FB6139  with expiration dates: 2015-07-14  2015-07-30  2015-08-21  2015-09-15  2015-10-02  2015-10-20  2015-11-11  2015-11-12  2015-12-08  2015-12-17  2015-12-18  2016-01-06  2016-01-26  2016-01-27  2016-02-16  2016-03-16  2016-04-06  2016-04-20  2016-04-21  2016-05-20    Dimension Vista CREA Lot Numbers:  14177BB  14212BD  14234BH  14261BC  14280BA  14301BB  14321AC  14346AC  15012AA  15042BA  15054AB  15076BA  15091BA  15125BB  15125BA  15141AA  with expiration dates: 2015-06-26  2015-07-31  2015-08-22  2015-09-18  2015-10-07  2015-10-28  2015-11-17  2015-12-12  2016-01-12  2016-02-11  2016-02-23  2016-03-16  2016-03-31  2016-05-04  2016-05-04  2016-05-22",
      "center_classification_date": "20150905",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and to the countries of : Afghanistan, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia Herzegovinia, Bulgaria, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Great Britain, Greece, Guatemala, Hungary, Iraq, Ireland, Honduras, Hong Kong, Iceland, Indonesia, India, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Libya, Macedonia, Malaysia, Mali, Mexico, Myanmar, Morocco, Netherlands, Nepal, Nicaragua, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkmenistan, United Arab Emirates,  Uruguay, Venezuela, Vietnam and Yemen.",
      "state": "DE",
      "product_description": "Dimension¿ CREA (Creatinine) and  Dimension Vista¿ CREA (Creatinine); The CREA method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended for the quantitative determination of creatinine in human serum, plasma and urine.",
      "report_date": "20150916",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2741-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71725",
      "termination_date": "20170324",
      "recall_initiation_date": "20150715",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}