  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Houston",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71936",
      "recalling_firm": "Trilliant Surgical Ltd.",
      "address_1": "6721 Portwest Dr Ste 160",
      "address_2": "N/A",
      "postal_code": "77024-8019",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX",
      "recall_number": "Z-2738-2015",
      "product_description": "2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE.  For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.",
      "product_quantity": "108 units",
      "reason_for_recall": "Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively.  Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.",
      "recall_initiation_date": "20150731",
      "center_classification_date": "20150904",
      "termination_date": "20170126",
      "report_date": "20150916",
      "code_info": "TSL002605"
    }
  ]
}