  "results": [
    {
      "status": "Terminated",
      "city": "Round Lake",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69122",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "25212 W. Illinois Route 120",
      "address_2": "N/A",
      "postal_code": "60073-9799",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution -- USA and Canada.",
      "recall_number": "Z-2738-2014",
      "product_description": "SPECTRUM Pump, Model No. 35700BAX.     Intended to be used for the controlled administration of intravenous fluids.",
      "product_quantity": "USA: 56 units, Canada: 2 units",
      "reason_for_recall": "One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.",
      "recall_initiation_date": "20140903",
      "center_classification_date": "20140930",
      "termination_date": "20170316",
      "report_date": "20141008",
      "code_info": "Software Versions 5.02.06, 6.02.06, and 6.02.11; Affected Serial Numbers: 712090, 723687, 723842, 724966, 725820, 735977, 751130, 752124, 755174, 768538, 771990, 774743, 781406, 783736, 784698, 794466, 805797, 808901, 809113, 812462, 815932, 842120, 857926, 858672, 862085, 863721, 865630, 873459, 889597, 890260, 903748, 912619, 920574, 920589, 956346, 957394, 965402, 966684, 976931, 977550, 978361, 983979, 984066, 984129, 984475,  985946, 987538, 993445, 995291, 996014, 996389, 1013037, 1004377,  1014565, 1014962, 967887, 950671, and 938428."
    }
  ]
}