  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bella Vista",
      "state": "N/A",
      "country": "Australia",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99304",
      "recalling_firm": "ResMed Ltd.",
      "address_1": "1 Elizabeth Macarthur Dr",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.",
      "recall_number": "Z-2735-2026",
      "product_description": "Resmed Astral 100/150    Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.",
      "product_quantity": "168,069 devices globally",
      "reason_for_recall": "Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.",
      "recall_initiation_date": "20260625",
      "center_classification_date": "20260728",
      "report_date": "20260805",
      "code_info": "Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070  Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070"
    }
  ]
}