{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient.  The issue was caused by a software defect in the workflow setup script.",
      "address_2": "",
      "product_quantity": "117 units",
      "code_info": "11000003, 11000008, 11000030, 11000004, 11000005, 11000006, 11000007, 11000009, 11000010, 11000011, 11000012, 11000013, 11000014, 11000016, 11000017, 11000018, 11000019, 11000020, 11000022, 11000023, 11000024, 11000025, 11000026, 11000027, 11000028, 11000029, 11000031, 11000032, 11000033, 11000034, 11000035, 11000036, 11000037, 11000038, 11000039, 11000040, 11000041, 11000042, 11000043, 11000044, 11000045, 11000046, 11000047, 11000048, 11000049, 11000050, 11000051, 11000052, 11000054, 11000055, 11000056, 11000057, 11000058, 11000059, 11000060, 11000061, 11000062, 11000063, 11000064, 11000065, 11000066, 11000067, 11000068, 11000069, 11000070, 11000071, 11000072, 11000073, 11000074, 11000075, 11000076, 11000077, 11000078, 11000079, 11000080, 11000081, 11000083, 11000084, 11000085, 11000086, 11000087, 11000088, 11000089, 11000090, 11000091, 11000092, 11000093, 11000094, 11000095, 11000096, 11000097, 11000098, 11000099, 11000100, 11000101, 11000102, 11000103, 11000104, 11000105, 11000106, 11000107, 11000108, 11000109, 11000110, 11000112, 11000113, 11000114, 11000115, 11000116, 11000117, 11000118, 11340001, 11340002, 11340003, 11340004, 11340005, 11000015",
      "center_classification_date": "20160903",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic of Lebanon, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.",
      "state": "OH",
      "product_description": "BrightView X, model #882478; gamma camera designed for single or dual detector nuclear imaging",
      "report_date": "20160914",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-2735-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74894",
      "termination_date": "20180702",
      "more_code_info": "",
      "recall_initiation_date": "20160525",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}