  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brooklyn Park",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72025",
      "recalling_firm": "Medtronic Perfusion Systems",
      "address_1": "7611 Northland Dr N",
      "address_2": "N/A",
      "postal_code": "55428-1088",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including the states of  AL, AR, CA, CO, FL, HI, ID, IN, KS, LA, MA, MS, MO, NY, NJ, OH, PA, TN, TX, UT, and WI, and the countries of Bolivia, Denmark, Dominican Republic, Italy, Latvia, Lebanon, Mexico, Panama, Singapore, Venezuela, and Zambia.",
      "recall_number": "Z-2735-2015",
      "product_description": "MEDIVATORS HEMOCOR HPH700, HPH-1400TS, HPH-1000TS.  Enables the perfusionist to manage the patient's hematocrit and fluid status.",
      "product_quantity": "1594 total (1306 US, 288 OUS)",
      "reason_for_recall": "Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification.  These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic Perfusion Tubing packs.",
      "recall_initiation_date": "20150818",
      "center_classification_date": "20150904",
      "termination_date": "20160509",
      "report_date": "20150916",
      "code_info": "759625A  759884A  759491A  762355A"
    }
  ]
}