  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69139",
      "recalling_firm": "Lumenis, Inc.",
      "address_1": "3959 W 1820 S",
      "address_2": "N/A",
      "postal_code": "84104",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to JAPAN, CHINA, and CANADA.",
      "recall_number": "Z-2733-2014",
      "product_description": "Array LaserLink, Manufactured by Lumenis,    The Array\" LaserLink\" is a laser system accessory intended for use in the treatment of ocular pathology. \" For the Posterior Segment, the Array\" LaserLink\" is indicated for use in Retinal Photocoagulation and Panretinal Photocoagulation of Vascular and Structural Abnormalities of the Retina and Choroid including: \" Proliferative and Severe and Very Severe Non-Proliferative Diabetic Retinopathy \" Macular Edema associated with Proliferative or Non-Proliferative Diabetic Retinopathy \" Choroidal Neovascularization \" Retinal Neovascularization associated with Retinal Occlusive Disease (Branch Retinal Vein Occlusion; Central Retinal Vein Occlusion) \" Macular Edema associated with Branch Retinal Vein Occlusion \" Retinal Tears and Detachments And Anterior Segments as follows: \" Iridotomy in Closed Angle Glaucoma \" Trabeculoplasty in Open Angle Glaucoma",
      "product_quantity": "16",
      "reason_for_recall": "Lumenis initiated a field-correction for the Array Laser Link&quot;, GA-0006700 (SN XXYYZZ) because the system may project an erroneous pattern display on the retina, which is different than the desired pattern.",
      "recall_initiation_date": "20140815",
      "center_classification_date": "20140930",
      "termination_date": "20141021",
      "report_date": "20141008",
      "code_info": "All assembled units since product release, Part Number: GA-0006700."
    }
  ]
}