  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mississauga",
      "state": "",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77534",
      "recalling_firm": "O-Two Medical Technologies, Inc.",
      "address_1": "7575 Kimbel Street",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Canada,  Belgium, Norway, and Malaysia",
      "recall_number": "Z-2731-2017",
      "product_description": "Burn Relief¿, MODEL #(s): 06BK1221, 06BK1232, 06BK1201, 06BK1200, CATALOG #(s): 06BK1221, 06BK1232, 06BK1201, 06BK1200 ---   DRESSING,WOUND,HYDROGEL W/OUT DRUG AND/OR BIOLOGIC. --- Burn Relief¿ products are self-cooling, non-adhering and are intended to be used on 1st and 2nd degree burns.",
      "product_quantity": "624 units",
      "reason_for_recall": "For lack of evidence to validate effectiveness of sterilization dose, O-Two Medical Technologies Inc. hereby announces a  voluntary medical device recall of the suspected batches of Burn Relief¿ products to prevent potential contamination.   .",
      "recall_initiation_date": "20150930",
      "center_classification_date": "20170713",
      "termination_date": "20170901",
      "report_date": "20170719",
      "code_info": "Lot/Batch#1150, 1152, 5210;     Expiration date: Jul.2018 (for lot 5210),  July 2020 (for lot 1150, 1152)",
      "more_code_info": ""
    }
  ]
}